Temperature-Modulated Differential Scanning Calorimeter
温度调制差示扫描量热仪
- SKU DSC-0003
- DSC
- CE (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | −90–600 °C |
|---|---|
| Modulation amplitude | ±0.01–2 °C |
| Modulation period | 10–100 s |
| Underlying heating rate | 0.1–5 °C/min |
| Temperature accuracy | ±0.1 °C |
| Heat capacity accuracy | ±2% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E967, ASTM E968, ASTM E1356
What is the Temperature-Modulated Differential Scanning Calorimeter?
A temperature-modulated DSC superimposes a small sinusoidal oscillation on the linear heating ramp, separating reversing from non-reversing heat flow. It resolves weak glass transitions that overlap melting and gives heat capacity in a single experiment.
What is the Temperature-Modulated Differential Scanning Calorimeter?
A temperature-modulated DSC superimposes a small sinusoidal oscillation on the linear heating ramp, separating reversing from non-reversing heat flow. It resolves weak glass transitions that overlap melting and gives heat capacity in a single experiment.
Overview
The furnace follows a linear ramp plus a periodic modulation of small amplitude and defined period, typically plus or minus 0.5 degrees Celsius with a period of 40 to 100 seconds. Fourier analysis of the resulting heat flow and heating rate signals separates the reversing component, related to heat capacity, from the non-reversing component, related to kinetic events such as crystallization, curing and enthalpy relaxation. In practice this resolves a glass transition hidden under a melting peak, separates curing from melting in thermoset systems and gives heat capacity without a separate sapphire run. The measurement requires calibration of heat capacity and of the modulation response, and the data treatment demands more care than conventional DSC, since the apparent signal depends on period, amplitude and total heating rate. Longer experiment times are the main penalty. The technique is used for pharmaceutical amorphous content, polymer blends, food systems and any sample with overlapping thermal events.
How does it work?
The sample and an empty reference pan sit on a sensor in a furnace whose temperature follows a program. In heat-flux instruments a single furnace heats both positions and thermocouples measure the small temperature difference that develops when the sample absorbs or releases heat; the difference is converted to heat flow using a calibration factor. Power-compensation instruments place the sample and reference in separate small furnaces and vary the heating power needed to keep both at the same temperature, recording power difference directly. Either way the result is a curve of heat flow against temperature or time, with the area under a peak giving transition enthalpy in joules per gram. A purge gas, usually nitrogen, prevents oxidation and removes volatiles, and cooling may use liquid nitrogen, a mechanical cooler or an air vortex. Temperature and enthalpy calibration use indium, zinc and other reference materials of known melting point and enthalpy.
What are the key features?
- Sinusoidal temperature modulation added to the linear ramp
- Separation of reversing heat capacity from non-reversing kinetic events
- Resolution of glass transitions overlapping melting peaks
- Heat capacity measured without a separate sapphire standard run
- Software with Fourier analysis and deconvolution routines
- Period and amplitude selectable for each sample type
What are the technical specifications?
A typical temperature-modulated differential scanning calorimeter configuration offers temperature range of −90–600 °C, modulation amplitude of ±0.01–2 °C, modulation period of 10–100 s, underlying heating rate of 0.1–5 °C/min, temperature accuracy of ±0.1 °C and heat capacity accuracy of ±2%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Temperature-Modulated Differential Scanning Calorimeter is used in amorphous content in pharmaceuticals, polymer blend compatibility studies, thermoset cure plus melting separation, food and starch transition analysis and heat capacity of thin films. Across the wider DSC class, laboratories also deploy it for polymer melting and crystallinity testing, pharmaceutical polymorph and purity screening and food starch and fat behaviour studies.
Which standards and certifications apply?
Relevant standards and methods include ASTM E967, ASTM E968, ASTM E1356, ASTM E1269, ISO 11357-1, ISO 11357-2 and USP <891>. Certification candidates for this model are CE, IEC 61010-1 safety and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Calibrate temperature and enthalpy before a measurement campaign, and verify at least one check standard afterwards. Choose the pan material and type for the temperature and chemistry involved: sealed or hermetic pans for volatile samples, high-pressure crucibles for pressure work and open pans for evolved gases. Keep sample mass similar between runs, because thermal lag grows with mass and shifts peaks to higher temperature. Use a defined purge gas flow, dry the cell if moisture condenses, and run an empty baseline periodically. Consumables such as pans, lids and seals are single-use, and crucibles used for pressure work must be inspected for deformation. Follow ASTM E967, E968 or ISO 11357 for calibration and reporting, and record pan type, gas, rate and mass with each curve.
How do I select the right Temperature-Modulated Differential Scanning Calorimeter?
Choose a modulated instrument when overlapping thermal events must be separated or heat capacity must be obtained from a single run, as with amorphous-content determination. For simple melting points and crystallinity, conventional DSC is faster and cheaper. In an RFQ, specify modulation parameters, underlying heating rate, software deconvolution needs and calibration standards.
How does it compare with other DSC variants?
Start with the temperature range you need, because it decides the cooling system: studies above room temperature need only an air or mechanical cooler, while glass transitions well below zero require liquid nitrogen. Then choose the measuring principle. Heat-flux instruments are robust, accept larger samples and cover most materials work; power-compensation instruments give sharper peaks and better resolution for small transitions and fast heating rates; temperature-modulated instruments separate reversing from non-reversing events; high-pressure cells are needed for oxidation stability and reactions under pressure. Check calorimetric accuracy and repeatability, resolution of weak transitions, sample pan options including sealed and hermetic pans, purge gas control, autosampler availability and the calibration standards supplied. Specify the heating rates, sample mass and the standards the method must follow.
Full specifications
| Temperature range | −90–600 °C |
|---|---|
| Modulation amplitude | ±0.01–2 °C |
| Modulation period | 10–100 s |
| Underlying heating rate | 0.1–5 °C/min |
| Temperature accuracy | ±0.1 °C |
| Heat capacity accuracy | ±2% |
| Signal separation | Reversing and non-reversing components |
| Purge gas | Nitrogen, argon or helium |
| Sample mass | 5–20 mg typical |
| Data analysis | Fourier analysis and deconvolution software |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Amorphous content in pharmaceuticals
- Polymer blend compatibility studies
- Thermoset cure plus melting separation
- Food and starch transition analysis
- Heat capacity of thin films
Key features
- Sinusoidal temperature modulation added to the linear ramp
- Separation of reversing heat capacity from non-reversing kinetic events
- Resolution of glass transitions overlapping melting peaks
- Heat capacity measured without a separate sapphire standard run
- Software with Fourier analysis and deconvolution routines
- Period and amplitude selectable for each sample type
Frequently asked questions about Temperature-Modulated Differential Scanning Calorimeter
Heat capacity responds instantaneously to the small oscillation, so it appears in the reversing signal, while kinetic events such as melting or curing lag behind the modulation and appear in the non-reversing signal. Fourier analysis of the heat flow and rate signals performs that separation mathematically.
The oscillation must complete several cycles during each transition, so the underlying heating rate is usually kept between 1 and 3 degrees Celsius per minute. That makes a modulated scan three to ten times slower than a conventional measurement of the same temperature range.
Two calibrations are needed: conventional temperature and enthalpy calibration with indium, and heat capacity calibration with sapphire using the same modulation parameters as the samples. Repeat the heat capacity calibration whenever the period or amplitude settings change.
It measures heat flow as a function of temperature, so it detects melting and crystallization temperatures and enthalpies, glass transitions, specific heat capacity, curing, polymorphic transitions, oxidation induction time and decomposition onset when the sample is sealed.
Temperature and enthalpy are calibrated with reference materials of certified melting point and enthalpy, commonly indium at 156.6 degrees Celsius, plus zinc or tin for higher ranges. Specific heat capacity is calibrated separately with sapphire using ASTM E1269 or an equivalent procedure.
Below the glass transition the material is rigid with low heat capacity; above it, molecular mobility raises the heat capacity, so the curve steps upward. The midpoint of that step is reported as the glass transition temperature, and the result depends on heating rate and thermal history.