Specific Protein Analyzer (Nephelometric)
特定蛋白分析仪(散射比浊法)
- SKU DIA-0003
- Diagnostic Analyzers
- CE (to be confirmed)
- ISO 13485 quality system (to be confirmed)
- NMPA registration (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Method | Rate or endpoint nephelometry |
|---|---|
| Throughput | 60–240 tests/h |
| Sample volume | 2–20 µL |
| CRP measuring range | 0.2–400 mg/L |
| Light source | Laser diode or LED, 600–700 nm |
| Precision (CV) | ≤ 5% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 15189, IEC 61010-2-101, CLSI C46
What is the Specific Protein Analyzer (Nephelometric)?
A nephelometric specific protein analyzer measures proteins such as CRP, immunoglobulins, complement and albumin by detecting light scattered by antigen-antibody complexes. It is selected by laboratories needing sensitive, low-volume quantitation of plasma proteins and inflammation markers.
What is the Specific Protein Analyzer (Nephelometric)?
A nephelometric specific protein analyzer measures proteins such as CRP, immunoglobulins, complement and albumin by detecting light scattered by antigen-antibody complexes. It is selected by laboratories needing sensitive, low-volume quantitation of plasma proteins and inflammation markers.
Overview
In rate or endpoint nephelometry, a laser or LED light beam passes through a cuvette in which antibody reagent has mixed with the patient sample, and a detector at a defined angle measures the light scattered by the forming immune complexes. Scatter intensity is proportional to antigen concentration within the working range, and antigen-excess checks prevent falsely low results. Menus commonly include CRP and high-sensitivity CRP, IgG, IgA, IgM, C3, C4, transferrin, prealbumin, haptoglobin, alpha-1 antitrypsin and cystatin C. Sample volume is typically 2 to 20 microliters, and throughput ranges from 60 to 240 tests per hour. Automatic dilution extends the measuring range. Compared with photometric chemistry analyzers, nephelometers offer high sensitivity for low-concentration proteins.
How does it work?
Diagnostic analyzers combine a sample handling system, a reaction or sensing chamber and a detector specific to the measurand. Blood gas analyzers use ion-selective, Clark-type oxygen and Severinghaus-type carbon dioxide electrodes. HbA1c analyzers separate hemoglobin fractions by cation-exchange HPLC and detect them at 415 nm. Specific protein analyzers measure light scattering or turbidity produced when antibody-coated particles bind a protein, using nephelometry or immunoturbidimetry. Blood culture systems monitor carbon dioxide production by colorimetric or fluorescent sensors inside bottles to flag microbial growth. Identification and susceptibility systems measure turbidity and growth kinetics in panels containing substrates or antibiotics. Calibration with standards and periodic quality control ensure that results remain traceable and within acceptance limits.
What are the key features?
- Rate or endpoint nephelometry with laser or LED light source
- Antigen-excess detection to prevent falsely low results
- Automatic sample dilution to extend the measuring range
- Sample volume of only 2–20 µL per test
- Menu of CRP, immunoglobulins, complement and transport proteins
- Multi-point calibration with stored curves and quality control
What are the technical specifications?
A typical specific protein analyzer (nephelometric) configuration offers method of Rate or endpoint nephelometry, throughput of 60–240 tests/h, sample volume of 2–20 µL, CRP measuring range of 0.2–400 mg/L, light source of Laser diode or LED, 600–700 nm and precision (CV) of ≤ 5%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Specific Protein Analyzer (Nephelometric) is used in CRP and inflammation monitoring, immunoglobulin quantification, complement testing and renal and nutritional protein testing. Across the wider Diagnostic Analyzers class, laboratories also deploy it for critical care blood gas monitoring, diabetes control monitoring and inflammation and immune protein testing.
Which standards and certifications apply?
Relevant standards and methods include ISO 15189, IEC 61010-2-101, CLSI C46, CLSI EP15, ISO 18113-1, ISO 15197, CLSI M47 and CLSI M07. Certification candidates for this model are CE, ISO 13485 quality system and NMPA registration; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install analyzers on stable benches with controlled temperature, stable power and adequate ventilation. Follow manufacturer schedules for electrode or column replacement, reagent pack changes and waste handling, and treat all specimens and waste as potentially infectious. Run quality control at defined intervals, participate in external quality assessment and verify performance on installation according to CLSI EP15. Blood gas analyzers require regular membrane and electrode maintenance and proper sample handling to avoid air bubbles. Blood culture instruments need a clean, vibration-free environment. Document calibration, maintenance and training under the laboratory quality system in line with ISO 15189, and plan service and spare parts support.
How do I select the right Specific Protein Analyzer (Nephelometric)?
Select a nephelometric analyzer when sensitive protein testing at low sample volume is priority. Define the protein menu, daily tests, required CRP range, dilution needs, reagent format, calibration traceability such as ERM-DA470k, and LIS connection in the RFQ.
How does it compare with other Diagnostic Analyzers variants?
Choose the analyzer by the measurand and the clinical setting. Blood gas analyzers for critical care should provide results in under two minutes from small sample volumes, with integrated co-oximetry and electrolytes if needed. For diabetes programs, HbA1c analyzers should demonstrate traceability to IFCC and NGSP and throughput matching the clinic. Specific protein analyzers differ in menu, sample dilution and calibration stability. Microbiology systems should match bottle capacity to blood culture volume. Compare reagent and consumable availability, maintenance demands, onboard calibration, LIS connectivity and operator training. Request evidence such as precision, interference testing and method comparison, and confirm the registration status of reagents in your market.
Full specifications
| Method | Rate or endpoint nephelometry |
|---|---|
| Throughput | 60–240 tests/h |
| Sample volume | 2–20 µL |
| CRP measuring range | 0.2–400 mg/L |
| Light source | Laser diode or LED, 600–700 nm |
| Precision (CV) | ≤ 5% |
| Reagent positions | 20–60 positions, refrigerated 2–10 °C |
| Sample positions | 40–100 positions |
| Dilution | Automatic, up to 1:1,000 |
| Power supply | 220 V, 50/60 Hz |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- CRP and inflammation monitoring
- Immunoglobulin quantification
- Complement testing
- Renal and nutritional protein testing
Key features
- Rate or endpoint nephelometry with laser or LED light source
- Antigen-excess detection to prevent falsely low results
- Automatic sample dilution to extend the measuring range
- Sample volume of only 2–20 µL per test
- Menu of CRP, immunoglobulins, complement and transport proteins
- Multi-point calibration with stored curves and quality control
Frequently asked questions about Specific Protein Analyzer (Nephelometric)
Nephelometry measures light scattered at an angle, typically 13 to 90 degrees, by immune complexes, while turbidimetry measures the reduction of transmitted light. Nephelometry is generally more sensitive at low concentrations, whereas turbidimetry runs on standard photometric analyzers.
At very high antigen concentrations, immune complexes dissolve and scatter decreases, which can produce a falsely low result. Analyzers detect this by adding a second antigen aliquot or checking kinetic patterns and then automatically dilute the sample and repeat.
Yes, analyzers with hs-CRP reagents and suitable calibration measure down to about 0.2 mg/L and are used for cardiovascular risk assessment alongside routine inflammation monitoring. Check the lower limit of quantitation and the reagent's claims before purchase.
They measure defined groups of markers such as blood gases and electrolytes, HbA1c, specific plasma proteins, or microbial growth and susceptibility. Each analyzer type uses the detection method suited to its measurand, such as electrodes, HPLC, light scattering or growth sensors.
Most modern blood gas analyzers produce pH, pCO2 and pO2 results in 30 to 90 seconds from a sample of 50 to 100 microliters, and may add electrolytes, glucose and lactate. Speed matters because arterial samples change with time and exposure to air.
HbA1c results guide diabetes diagnosis and treatment, so different methods must agree. Analyzers should be traceable to IFCC reference methods and certified to NGSP criteria, so that a result of 6.5 percent means the same across laboratories and over time.