Fiber-Optic CCD UV-Vis Spectrometer
光纤 CCD 紫外可见光谱仪
- SKU UVV-0002
- UV-Vis
- CE (to be confirmed)
- RoHS (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Wavelength range | 200–850 nm or 350–1100 nm |
|---|---|
| Detector | CCD, 2048 pixels |
| Optical resolution (FWHM) | 0.3–2.0 nm |
| Integration time | 1 ms–10 s |
| Signal-to-noise ratio | ≥250:1 (single acquisition) |
| Stray light | ≤0.1% |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E275, ASTM E925, ASTM E958
What is the Fiber-Optic CCD UV-Vis Spectrometer?
A fiber-optic CCD UV-Vis spectrometer is a compact modular instrument that couples light from a source, cuvette holder or probe via optical fiber to a CCD array. It is the practical choice for in-situ, process and OEM measurements, with a spectrum available in milliseconds.
What is the Fiber-Optic CCD UV-Vis Spectrometer?
A fiber-optic CCD UV-Vis spectrometer is a compact modular instrument that couples light from a source, cuvette holder or probe via optical fiber to a CCD array. It is the practical choice for in-situ, process and OEM measurements, with a spectrum available in milliseconds.
Overview
The fiber-optic spectrometer houses a fixed grating, entrance slit and back-thinned or linear CCD detector in a palm-size bench. Light reaches it through a 200–600 µm SMA-terminated fiber, so the light source, sampling accessory and spectrometer can be placed separately. Options include cuvette holders, dip probes, flow cells, reflectance probes and integrating spheres. Integration times span 1 ms to several seconds, and USB-powered operation makes the system portable. Optical resolution of 0.3–2 nm is achieved depending on slit and grating, with spectral ranges typically 200–850 nm or 350–1100 nm for a single unit. Dynamic range and stray light are lower than laboratory-grade scanning instruments, about 0.1% stray light, but flexibility is high. Dark-current subtraction and lamp referencing are essential for reliable quantification.
How does it work?
UV-Vis instruments pass light of selected wavelength through or onto a sample and measure the ratio of transmitted or reflected intensity to a reference. Sources are typically a deuterium lamp for ultraviolet and a tungsten-halogen lamp for visible and near-infrared, often replaced by xenon flash lamps or LEDs in compact designs. Wavelength selection uses a scanning monochromator with a holographic grating, or a fixed grating that disperses light onto a diode-array or CCD detector so the whole spectrum is captured at once. Photomultiplier tubes, silicon photodiodes or InGaAs detectors convert the light to an electrical signal. Absorbance is calculated as −log10 of transmittance and relates to concentration by the Beer–Lambert law. Stray light, bandwidth and photometric linearity define performance at high absorbance, while integrating spheres and fiber probes adapt the same principle to solids, films and in-situ measurements.
What are the key features?
- Palm-size CCD spectrometer operates from USB power for field and process use.
- Fiber coupling allows the source, sample and detector to be placed separately.
- Integration time from 1 ms to several seconds for varying signal levels.
- Dip probes, flow cells and reflectance probes enable in-situ measurements.
- Configurable gratings and slits trade resolution against sensitivity.
- SDK and API support OEM integration and custom applications.
What are the technical specifications?
A typical fiber-optic ccd uv-vis spectrometer configuration offers wavelength range of 200–850 nm or 350–1100 nm, detector of CCD, 2048 pixels, optical resolution (FWHM) of 0.3–2.0 nm, integration time of 1 ms–10 s, signal-to-noise ratio of ≥250:1 (single acquisition) and stray light of ≤0.1%. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Fiber-Optic CCD UV-Vis Spectrometer is used in in-situ process monitoring, reflectance and color measurement, plasma and LED spectral analysis, field water quality measurements and educational and OEM instruments. Across the wider UV-Vis class, laboratories also deploy it for reaction kinetics and enzyme assays, nanoparticle plasmon resonance characterization and thin-film and coating optical properties.
Which standards and certifications apply?
Relevant standards and methods include ASTM E275, ASTM E925, ASTM E958, USP <857>, Ph. Eur. 2.2.25, ISO 8655-7 and GB/T 26798. Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Allow lamps to warm up for 15–30 minutes and perform a baseline correction before every measurement session. Verify performance periodically using holmium oxide or didymium filters for wavelength accuracy, potassium dichromate solutions for photometric accuracy and sodium iodide or potassium chloride solutions for stray light. Use matched quartz cuvettes for the UV and keep optical surfaces clean and free of fingerprints. Replace the deuterium lamp after its rated life of around 2000 hours. Control ambient temperature and vibration, and keep dust and solvent vapors away from the optics. For regulated laboratories, schedule IQ/OQ/PQ and use software with audit trails and access control.
How do I select the right Fiber-Optic CCD UV-Vis Spectrometer?
Select a fiber-optic CCD spectrometer for modular, remote or embedded measurements where footprint and flexibility matter more than ultimate photometric range. For high-absorbance laboratory quantification, a benchtop scanning instrument is better. In an RFQ, specify range, resolution, fiber diameter, accessories, light source and software interface.
How does it compare with other UV-Vis variants?
Choose a UV-Vis spectroscopy instrument according to spectral speed, wavelength range, sample type and accessory needs. Diode-array and CCD instruments capture a full spectrum in under a second and suit kinetics or unstable samples, but stray light is higher than in scanning double-beam systems that remain the choice for high-absorbance work. Extending to 3300 nm with a PbS or InGaAs detector requires a UV-Vis-NIR system. Solid films, powders and coatings need an integrating sphere. Modular fiber-optic spectrometers fit process lines and OEM use, while microplate readers serve high-throughput assays. Compare bandwidth, stray light at 220 and 340 nm, photometric range, baseline flatness and software features. Specify temperature control if thermodynamic or enzyme kinetic measurements are planned.
Full specifications
| Wavelength range | 200–850 nm or 350–1100 nm |
|---|---|
| Detector | CCD, 2048 pixels |
| Optical resolution (FWHM) | 0.3–2.0 nm |
| Integration time | 1 ms–10 s |
| Signal-to-noise ratio | ≥250:1 (single acquisition) |
| Stray light | ≤0.1% |
| Fiber connector | SMA 905 |
| Dynamic range | ≥1500:1 (single acquisition) |
| Interface | USB 2.0/3.0 |
| Dimensions | ≈90 × 65 × 35 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- In-situ process monitoring
- Reflectance and color measurement
- Plasma and LED spectral analysis
- Field water quality measurements
- Educational and OEM instruments
Key features
- Palm-size CCD spectrometer operates from USB power for field and process use.
- Fiber coupling allows the source, sample and detector to be placed separately.
- Integration time from 1 ms to several seconds for varying signal levels.
- Dip probes, flow cells and reflectance probes enable in-situ measurements.
- Configurable gratings and slits trade resolution against sensitivity.
- SDK and API support OEM integration and custom applications.
Frequently asked questions about Fiber-Optic CCD UV-Vis Spectrometer
Typical optical resolution is 0.3–2 nm FWHM, depending on slit width and grating groove density. Narrower slits improve resolution but reduce light throughput, so integration time must increase to maintain signal-to-noise.
Yes. With a cuvette holder, a deuterium-halogen or LED source and fibers, absorbance is computed from a dark and reference spectrum. Limited stray-light performance restricts reliable quantification to approximately 0.1–2 A.
A CCD accumulates thermal signal even without light, especially at long integration times and high temperature. Subtracting a dark spectrum taken with the same settings removes this background and prevents baseline offsets in absorbance and reflectance calculations.
A diode-array instrument disperses all wavelengths onto a detector array and records the whole spectrum in under a second, which suits kinetics. A scanning system steps a monochromator through wavelengths one at a time, giving lower stray light and wider photometric range for demanding samples.
Standard UV-Vis instruments cover roughly 190–1100 nm, spanning deep ultraviolet to short-wave near infrared. UV-Vis-NIR versions extend to 2500 or 3300 nm using PbS or InGaAs detectors, which is needed for glass, optical coatings and some polymers.
Most instruments give best linearity between about 0.1 and 1.0 absorbance units, and high-quality double-beam systems remain linear up to 3 A or more. Above 2 A, stray light can cause deviations, so dilute samples or use shorter path cuvettes when possible.