Diode-Array UV-Vis Spectrophotometer
二极管阵列紫外可见分光光度计
- SKU UVV-0001
- UV-Vis
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Wavelength range | 190–1100 nm |
|---|---|
| Detector | Photodiode array, 1024 elements |
| Spectral bandwidth | 1.5 nm |
| Wavelength accuracy | ±0.5 nm |
| Photometric range | −0.3 to 3.0 A |
| Photometric accuracy | ±0.005 A at 1 A |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E275, ASTM E925, ASTM E958
What is the Diode-Array UV-Vis Spectrophotometer?
A diode-array UV-Vis spectrophotometer images a full 190–1100 nm spectrum onto a detector array in well under one second. It is the appropriate choice for kinetic studies, unstable samples and chromatographic peak purity work where simultaneous multi-wavelength information is needed.
What is the Diode-Array UV-Vis Spectrophotometer?
A diode-array UV-Vis spectrophotometer images a full 190–1100 nm spectrum onto a detector array in well under one second. It is the appropriate choice for kinetic studies, unstable samples and chromatographic peak purity work where simultaneous multi-wavelength information is needed.
Overview
In the diode-array design, polychromatic light passes through the sample first and is then dispersed by a fixed holographic grating onto a photodiode array of 512 to 2048 elements, so no moving parts are involved in wavelength selection. A full spectrum is acquired in 0.1–1 second, and multiple spectra can be collected in rapid succession to follow fast reactions. Typical resolution is 1–2 nm with wavelength accuracy of ±0.5 nm, and a xenon flash lamp gives long lamp life with low heat load. Stray light is higher than in a scanning double-beam instrument, often below 0.05% T at 220 nm, which limits reliable absorbance to roughly 2 A. Software tools include multi-component analysis, derivative spectra and peak purity. The compact footprint suits benchtop use and integration with HPLC as a diode-array detector.
How does it work?
UV-Vis instruments pass light of selected wavelength through or onto a sample and measure the ratio of transmitted or reflected intensity to a reference. Sources are typically a deuterium lamp for ultraviolet and a tungsten-halogen lamp for visible and near-infrared, often replaced by xenon flash lamps or LEDs in compact designs. Wavelength selection uses a scanning monochromator with a holographic grating, or a fixed grating that disperses light onto a diode-array or CCD detector so the whole spectrum is captured at once. Photomultiplier tubes, silicon photodiodes or InGaAs detectors convert the light to an electrical signal. Absorbance is calculated as −log10 of transmittance and relates to concentration by the Beer–Lambert law. Stray light, bandwidth and photometric linearity define performance at high absorbance, while integrating spheres and fiber probes adapt the same principle to solids, films and in-situ measurements.
What are the key features?
- Full spectrum acquired in 0.1–1 second without moving wavelength-selection parts.
- Xenon flash lamp offers long life, low heat load and no warm-up wait.
- Multi-component analysis resolves overlapping absorbers with least-squares fitting.
- Derivative and spectral math tools support peak purity and baseline handling.
- Compact benchtop layout with optional multi-cell and Peltier accessories.
- Wavelength verification routines built in with holmium filter.
What are the technical specifications?
A typical diode-array uv-vis spectrophotometer configuration offers wavelength range of 190–1100 nm, detector of Photodiode array, 1024 elements, spectral bandwidth of 1.5 nm, wavelength accuracy of ±0.5 nm, photometric range of −0.3 to 3.0 A and photometric accuracy of ±0.005 A at 1 A. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Diode-Array UV-Vis Spectrophotometer is used in reaction and enzyme kinetics, HPLC peak purity verification, dye and pigment analysis, quality control of colored solutions and teaching laboratories. Across the wider UV-Vis class, laboratories also deploy it for reaction kinetics and enzyme assays, nanoparticle plasmon resonance characterization and thin-film and coating optical properties.
Which standards and certifications apply?
Relevant standards and methods include ASTM E275, ASTM E925, ASTM E958, USP <857>, Ph. Eur. 2.2.25, ISO 8655-7 and GB/T 26798. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Allow lamps to warm up for 15–30 minutes and perform a baseline correction before every measurement session. Verify performance periodically using holmium oxide or didymium filters for wavelength accuracy, potassium dichromate solutions for photometric accuracy and sodium iodide or potassium chloride solutions for stray light. Use matched quartz cuvettes for the UV and keep optical surfaces clean and free of fingerprints. Replace the deuterium lamp after its rated life of around 2000 hours. Control ambient temperature and vibration, and keep dust and solvent vapors away from the optics. For regulated laboratories, schedule IQ/OQ/PQ and use software with audit trails and access control.
How do I select the right Diode-Array UV-Vis Spectrophotometer?
Select the diode-array model for speed, reliability and multi-wavelength capture, for instance in kinetics and multi-component quantification. It is not the first choice for samples above 2 A. In an RFQ, specify wavelength range, resolution, stray light, cell holder options, temperature control and software features.
How does it compare with other UV-Vis variants?
Choose a UV-Vis spectroscopy instrument according to spectral speed, wavelength range, sample type and accessory needs. Diode-array and CCD instruments capture a full spectrum in under a second and suit kinetics or unstable samples, but stray light is higher than in scanning double-beam systems that remain the choice for high-absorbance work. Extending to 3300 nm with a PbS or InGaAs detector requires a UV-Vis-NIR system. Solid films, powders and coatings need an integrating sphere. Modular fiber-optic spectrometers fit process lines and OEM use, while microplate readers serve high-throughput assays. Compare bandwidth, stray light at 220 and 340 nm, photometric range, baseline flatness and software features. Specify temperature control if thermodynamic or enzyme kinetic measurements are planned.
Full specifications
| Wavelength range | 190–1100 nm |
|---|---|
| Detector | Photodiode array, 1024 elements |
| Spectral bandwidth | 1.5 nm |
| Wavelength accuracy | ±0.5 nm |
| Photometric range | −0.3 to 3.0 A |
| Photometric accuracy | ±0.005 A at 1 A |
| Stray light | ≤0.05% T at 220 nm (NaI) |
| Acquisition time (full spectrum) | ≤1 s |
| Light source | Xenon flash lamp |
| Interfaces | USB, Ethernet |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Reaction and enzyme kinetics
- HPLC peak purity verification
- Dye and pigment analysis
- Quality control of colored solutions
- Teaching laboratories
Key features
- Full spectrum acquired in 0.1–1 second without moving wavelength-selection parts.
- Xenon flash lamp offers long life, low heat load and no warm-up wait.
- Multi-component analysis resolves overlapping absorbers with least-squares fitting.
- Derivative and spectral math tools support peak purity and baseline handling.
- Compact benchtop layout with optional multi-cell and Peltier accessories.
- Wavelength verification routines built in with holmium filter.
Frequently asked questions about Diode-Array UV-Vis Spectrophotometer
Yes, for most routine work in the 0.1–2 A range. Photometric accuracy of ±0.005 A is typical. For absorbance above about 2 A or very low stray-light requirements, a scanning double-beam instrument gives better performance.
Yes. The sample sees the full polychromatic beam, which can photodegrade light-sensitive compounds. A shutter limits exposure to the measurement, and xenon flash lamps reduce heating. For very photosensitive samples, a scanning instrument with monochromatic illumination is preferable.
Yes, this is a primary use. It can record complete spectra every 0.1 s, track several wavelengths simultaneously and fit rate constants, and works with a Peltier cell holder and magnetic stirrer for temperature-controlled measurements.
A diode-array instrument disperses all wavelengths onto a detector array and records the whole spectrum in under a second, which suits kinetics. A scanning system steps a monochromator through wavelengths one at a time, giving lower stray light and wider photometric range for demanding samples.
Standard UV-Vis instruments cover roughly 190–1100 nm, spanning deep ultraviolet to short-wave near infrared. UV-Vis-NIR versions extend to 2500 or 3300 nm using PbS or InGaAs detectors, which is needed for glass, optical coatings and some polymers.
Most instruments give best linearity between about 0.1 and 1.0 absorbance units, and high-quality double-beam systems remain linear up to 3 A or more. Above 2 A, stray light can cause deviations, so dilute samples or use shorter path cuvettes when possible.
Related models



UV-Vis Spectrophotometer with Peltier Multi-Cell Kinetics System
带帕尔贴多池动力学系统的紫外可见分光光度计
- CE (to be confirmed)
- +1