Skip to main content
Lab Manufacturers China
Ultracentrifuge with Vertical and Near-Vertical Rotors — product photo

Ultracentrifuge with Vertical and Near-Vertical Rotors

配垂直和近垂直转子的超速离心机

  • SKU ULT-0005
  • Ultracentrifuges
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-2-020 compliance (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Maximum speed50,000–100,000 rpm
Maximum RCF200,000–500,000 x g
Rotor angle0° (vertical) or 7–10° (near-vertical)
Rotor capacity8 x 5 mL to 8 x 39 mL
Typical isopycnic run time1–4 h
Temperature range0 to 40 °C
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-020, IEC 61326-1

What is the Ultracentrifuge with Vertical and Near-Vertical Rotors?

An ultracentrifuge fitted with vertical and near-vertical rotors holds tubes upright or at 7 to 10 degrees, shortening the path length so isopycnic bands form in 1 to 4 hours instead of 16 to 48. It suits plasmid, virus and lipoprotein banding in cesium chloride or iodixanol gradients.

What is the Ultracentrifuge with Vertical and Near-Vertical Rotors?

An ultracentrifuge fitted with vertical and near-vertical rotors holds tubes upright or at 7 to 10 degrees, shortening the path length so isopycnic bands form in 1 to 4 hours instead of 16 to 48. It suits plasmid, virus and lipoprotein banding in cesium chloride or iodixanol gradients.

Overview

Vertical rotors keep tubes parallel to the rotation axis, so during the spin the gradient reorients by 90 degrees and particles travel only the tube diameter, typically 1 cm, to reach their isopycnic position. Near-vertical rotors tilt tubes by 7 to 10 degrees and give a similar shortened path with less risk of smearing. Run times of 1 to 4 hours replace the 16 to 48 hours needed in fixed-angle or swing-bucket rotors. Typical capacities are 8 x 5 mL to 8 x 39 mL, with speeds of 50,000 to 100,000 rpm and RCF of 200,000 to 500,000 x g. The resolution is slightly lower than in swing-bucket rotors but adequate for preparative banding. Reorienting gradients require slow acceleration and deceleration ramps, which the instrument offers as programmable profiles. Sealed tubes are normally required.

How does it work?

An ultracentrifuge drives a titanium or carbon fiber rotor through a high-frequency induction or brushless motor, typically with a magnetically suspended or oil-lubricated flexible shaft. The chamber is evacuated below 0.1 Pa (1 micron Hg) by a rotary vane and diffusion or turbomolecular pump, eliminating air friction that would otherwise overheat the rotor. Refrigeration controls rotor temperature via infrared sensing, usually between 0 and 40 C. Particles sediment according to Stokes' law, with sedimentation coefficients measured in Svedberg units, in preparative runs, or are separated by buoyant density in cesium chloride, sucrose or iodixanol gradients in isopycnic runs. Rotors are identified automatically, and the control system tracks run hours and cycles to derate them. A heavy armor ring contains rotor fragments in the event of failure.

What are the key features?

  • Vertical tube orientation shortens the path to about one tube diameter
  • Isopycnic runs complete in 1 to 4 hours instead of 16 to 48 hours
  • Near-vertical rotors at 7 to 10 degrees reduce band smearing
  • Capacities from 8 x 5 mL to 8 x 39 mL per rotor
  • Programmable slow ramps control gradient reorientation during acceleration

What are the technical specifications?

A typical ultracentrifuge with vertical and near-vertical rotors configuration offers maximum speed of 50,000–100,000 rpm, maximum RCF of 200,000–500,000 x g, rotor angle of 0° (vertical) or 7–10° (near-vertical), rotor capacity of 8 x 5 mL to 8 x 39 mL, typical isopycnic run time of 1–4 h and temperature range of 0 to 40 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Ultracentrifuge with Vertical and Near-Vertical Rotors is used in plasmid DNA banding in cesium chloride, virus and viral vector banding, lipoprotein fractionation and rapid isopycnic gradient separations. Across the wider Ultracentrifuges class, laboratories also deploy it for virus and viral vector purification, exosome and extracellular vesicle isolation and lipoprotein fractionation.

Which standards and certifications apply?

Relevant standards and methods include IEC 61010-1, IEC 61010-2-020, IEC 61326-1, ISO 9001, EN 12469 and GB 4793.1. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-2-020 compliance; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Floor-standing units weigh 200 to 450 kg and often require a dedicated 220 V or 380 V circuit and a level, vibration-free floor. Rotor management is critical: log every run, derate titanium and aluminum rotors per the maker's schedule, and retire rotors at the stated lifetime. Use only tubes and adapters certified for the rotor, and fill them as specified. Vacuum pumps need periodic oil changes, and seals require inspection. Clean rotors with neutral detergent and never use chloride or strong alkali on aluminum. Annual preventive maintenance should include speed, temperature and vacuum checks. Install and operate in line with IEC 61010-2-020 and local pressure-vessel and biosafety rules.

How do I select the right Ultracentrifuge with Vertical and Near-Vertical Rotors?

Choose vertical or near-vertical rotors when run time matters and banding resolution requirements are moderate. They are not suitable for pelleting or step gradients needing maximum resolution, where swing-bucket siblings are better. In the RFQ, specify tube sizes, gradient media, RCF and whether sealed tubes and sealing equipment are needed.

How does it compare with other Ultracentrifuges variants?

Specify the maximum RCF needed: 100,000 to 150,000 x g suits most gradient and pelleting work, while ribosomes, small viruses and proteins need 300,000 to 600,000 x g or analytical capability. Choose rotor geometry to match the task, with swing-bucket rotors for gradients, fixed-angle rotors for pelleting, vertical or near-vertical rotors for fast gradients and zonal or continuous-flow rotors for large volumes. Consider tube materials and sealing, because ultracentrifuge tubes require heat-sealed or cap-assembled closures. Evaluate run documentation, rotor life tracking, temperature range, vacuum time, noise, and installation needs including weight, power and cooling water. For multi-user facilities, autoclavable rotors and biosafety containment options matter.

Full specifications

Full specifications
Maximum speed50,000–100,000 rpm
Maximum RCF200,000–500,000 x g
Rotor angle0° (vertical) or 7–10° (near-vertical)
Rotor capacity8 x 5 mL to 8 x 39 mL
Typical isopycnic run time1–4 h
Temperature range0 to 40 °C
Tube typeSealed polyallomer or ultra-clear
Acceleration profilesSlow ramp to 3,000 rpm then full
Noise level≤ 60 dB(A)
Power supply220 V, 50/60 Hz, 3–5 kW

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Plasmid DNA banding in cesium chloride
  • Virus and viral vector banding
  • Lipoprotein fractionation
  • Rapid isopycnic gradient separations

Key features

  • Vertical tube orientation shortens the path to about one tube diameter
  • Isopycnic runs complete in 1 to 4 hours instead of 16 to 48 hours
  • Near-vertical rotors at 7 to 10 degrees reduce band smearing
  • Capacities from 8 x 5 mL to 8 x 39 mL per rotor
  • Programmable slow ramps control gradient reorientation during acceleration

Frequently asked questions about Ultracentrifuge with Vertical and Near-Vertical Rotors

In a vertical rotor the sedimentation path equals the tube diameter, around 1 cm, rather than the full tube length of 5 to 10 cm. Particles reach their isopycnic position much sooner, cutting run times from 16 to 48 hours to 1 to 4 hours.

Related models

All Ultracentrifuges

Request a quotation for the Ultracentrifuge with Vertical and Near-Vertical Rotors

Send your configuration, quantity and destination port to sales@labmanufacturerschina.com; we reply with a specification sheet and price.

Request a quote