Refrigerated Microplate Incubator
制冷型微孔板培养箱
- SKU MII-0004
- Microplate Incubators
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | 4–60 °C |
|---|---|
| Temperature uniformity | ±0.5 °C |
| Cooling method | Peltier or compressor |
| Chamber volume | 20–100 L |
| Plate capacity | 4–40 plates |
| Program steps | ≥ 10 steps, time-based |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-010, IEC 61326-1
What is the Refrigerated Microplate Incubator?
A refrigerated microplate incubator controls temperature both below and above ambient, typically from 4 to 60 degrees Celsius, to hold plates for cold binding steps, enzyme kinetics and temperature-sensitive assays. It allows cold overnight incubations and warm steps in one unit.
What is the Refrigerated Microplate Incubator?
A refrigerated microplate incubator controls temperature both below and above ambient, typically from 4 to 60 degrees Celsius, to hold plates for cold binding steps, enzyme kinetics and temperature-sensitive assays. It allows cold overnight incubations and warm steps in one unit.
Overview
Refrigerated incubators combine Peltier or compressor cooling with heating elements and forced-air circulation, allowing set points from about 4 to 60 degrees Celsius with uniformity of plus or minus 0.5 degrees. Chamber volumes of 20 to 100 liters hold several plates, and some include shaking platforms and humidity control. Programmable temperature profiles change set points over hours, for example 37 degrees for an hour followed by 4 degrees overnight, without handling plates. Condensation is managed by air circulation, sealed plates and optionally a heated lid on the door. Compressor-based units have higher cooling capacity, but vibration should be checked for sensitive assays. Power use is higher than heated-only units. Cooling below dew point in humid rooms can form condensation inside, so seals and drains matter.
How does it work?
Microplate incubators maintain a uniform thermal environment around plates. Forced-air designs circulate heated air through the chamber with a fan, heater and a platinum resistance sensor feeding a PID controller, giving uniformity within about plus or minus 0.5 degrees Celsius. Dry-block designs conduct heat from an aluminum block that contacts the plate base, with Peltier elements for heating and cooling. Humidity is generated by a water reservoir or ultrasonic humidifier and monitored by a sensor, preventing evaporation from wells. CO2 incubators add infrared or thermal conductivity sensors and gas solenoid valves to hold 5 percent carbon dioxide for bicarbonate-buffered media. Shaking incubators combine these conditions with orbital motion. Door seals, insulation and heated lids limit gradients and condensation, which otherwise produce edge effects between outer and inner wells.
What are the key features?
- Temperature range from 4 to 60 degrees Celsius in one unit
- Peltier or compressor cooling with forced-air circulation
- Programmable profiles for warm-to-cold transitions
- Uniformity of plus or minus 0.5 degrees Celsius
- Optional shaking platform and humidity control
- Chamber volumes of 20 to 100 liters
What are the technical specifications?
A typical refrigerated microplate incubator configuration offers temperature range of 4–60 °C, temperature uniformity of ±0.5 °C, cooling method of Peltier or compressor, chamber volume of 20–100 L, plate capacity of 4–40 plates and program steps of ≥ 10 steps, time-based. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Refrigerated Microplate Incubator is used in cold overnight ELISA binding, enzyme kinetics at controlled temperatures, protein stability and aggregation assays, temperature-shift incubations and antibody and antigen coating at 4 °C. Across the wider Microplate Incubators class, laboratories also deploy it for ELISA and immunoassay incubation, bacterial growth and susceptibility testing and mammalian cell culture in plates.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-010, IEC 61326-1, ANSI/SLAS 1-2004 to 4-2004, ISO 13485 (IVD workflows) and EN 12469 (for biosafety-related enclosures, where applicable). Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install on a level surface away from drafts, direct sunlight and heat sources, with clearance for ventilation. Map temperature with calibrated probes in loaded and empty conditions at installation and periodically, and record set point drift. Use distilled or deionized water in humidity reservoirs and clean them regularly to prevent microbial growth. Clean interior surfaces with disinfectants compatible with the chamber material, and run sterilization cycles where available. Avoid frequent door openings, and allow recovery time for temperature and CO2. Calibrate CO2 sensors against a reference analyzer, and keep alarms active for unattended runs.
How do I select the right Refrigerated Microplate Incubator?
Choose this variant when assays need both cold and warm incubation steps or set points near or below ambient. For warm-only work, a simpler incubator is cheaper. In an RFQ specify temperature range, cooling method, uniformity, programmable profiles, humidity and plate capacity.
How does it compare with other Microplate Incubators variants?
Match the incubator to assay temperature, duration and atmosphere. Dry-block or forced-air models are enough for ELISA and binding steps at 25 to 37 degrees. Use a humidified chamber for long incubations to limit evaporation. Mammalian cell assays need CO2 and humidity control, while bacterial growth may need shaking and temperatures up to 60 degrees. Check capacity in plates, uniformity and recovery time after door opening, plus cooling below ambient for refrigerated incubations. Consider footprint, compatibility with stackers and readers, contamination control features such as HEPA filtration or copper surfaces, data logging and alarms. For regulated laboratories, verify temperature mapping documentation, calibration certificates and audit trail capabilities.
Full specifications
| Temperature range | 4–60 °C |
|---|---|
| Temperature uniformity | ±0.5 °C |
| Cooling method | Peltier or compressor |
| Chamber volume | 20–100 L |
| Plate capacity | 4–40 plates |
| Program steps | ≥ 10 steps, time-based |
| Cooling rate | ≈ 1 °C/min (37 to 4 °C) |
| Humidity control | Optional, 60–90% RH |
| Shaking (optional) | Orbital, 100–800 rpm |
| Alarms | Over/under temperature, door |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Cold overnight ELISA binding
- Enzyme kinetics at controlled temperatures
- Protein stability and aggregation assays
- Temperature-shift incubations
- Antibody and antigen coating at 4 °C
Key features
- Temperature range from 4 to 60 degrees Celsius in one unit
- Peltier or compressor cooling with forced-air circulation
- Programmable profiles for warm-to-cold transitions
- Uniformity of plus or minus 0.5 degrees Celsius
- Optional shaking platform and humidity control
- Chamber volumes of 20 to 100 liters
Frequently asked questions about Refrigerated Microplate Incubator
Peltier cooling is quiet, vibration-free and compact, suiting smaller chambers. Compressor cooling delivers higher capacity and faster pull-down for larger volumes but adds vibration and noise.
Use sealed plate lids or film, keep air circulation on, avoid opening the door in humid rooms and allow plates to reach room temperature slowly before opening seals to prevent water forming on the surfaces.
Yes. Most units store multi-step profiles with ramp times and hold times, for example a warm reaction followed by cold hold, so overnight processes run unattended.
Most ELISA steps are run at 37 degrees Celsius for 30 to 120 minutes, or at room temperature, or overnight at 2 to 8 degrees, depending on the kit. Incubators should hold the set temperature uniformly across all wells.
The edge effect is a difference in results between outer and inner wells caused by faster evaporation and temperature gradients at plate edges. Humidified chambers, sealing films and filling outer wells with liquid reduce it.
Only for mammalian cell assays in bicarbonate-buffered media, which require about 5 percent carbon dioxide to keep pH. ELISA, bacterial growth and HEPES-buffered assays do not need CO2.