Forced-Air Microplate Incubator with Humidity Control
带湿度控制的强制风微孔板培养箱
- SKU MII-0002
- Microplate Incubators
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- RoHS (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Temperature range | Ambient +5 to 60 °C |
|---|---|
| Temperature uniformity | ±0.5 °C |
| Humidity range | 60–95% RH |
| Chamber volume | 20–150 L |
| Plate capacity | 4–60 plates (rack dependent) |
| Recovery after door open | ≤ 10 min to 37 °C |
By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-010, IEC 61326-1
What is the Forced-Air Microplate Incubator with Humidity Control?
A forced-air microplate incubator circulates heated, humidified air around stacked plates, holding 37 degrees Celsius with relative humidity above 80 percent for long ELISA and assay incubations. It reduces evaporation and edge effects across many plates.
What is the Forced-Air Microplate Incubator with Humidity Control?
A forced-air microplate incubator circulates heated, humidified air around stacked plates, holding 37 degrees Celsius with relative humidity above 80 percent for long ELISA and assay incubations. It reduces evaporation and edge effects across many plates.
Overview
These incubators use a fan to move air through a chamber holding one to several dozen plates in racks, with heater, sensor and humidifier controlled by a microprocessor. Temperature ranges from ambient plus 5 to 60 degrees with uniformity of plus or minus 0.5 degrees, and humidity is regulated from about 60 to 95 percent relative humidity by a water reservoir or ultrasonic mist. Sealed plate lids and humidity reduce edge effects over 4 to 24 hour incubations. Shelf systems or racks with spacing allow airflow to each plate, and large models integrate with plate stackers and robotic arms. Chamber volume ranges from 20 to 150 liters. Recovery after door opening takes 5 to 15 minutes, and the humidity system needs clean water and regular sanitizing to avoid microbial growth.
How does it work?
Microplate incubators maintain a uniform thermal environment around plates. Forced-air designs circulate heated air through the chamber with a fan, heater and a platinum resistance sensor feeding a PID controller, giving uniformity within about plus or minus 0.5 degrees Celsius. Dry-block designs conduct heat from an aluminum block that contacts the plate base, with Peltier elements for heating and cooling. Humidity is generated by a water reservoir or ultrasonic humidifier and monitored by a sensor, preventing evaporation from wells. CO2 incubators add infrared or thermal conductivity sensors and gas solenoid valves to hold 5 percent carbon dioxide for bicarbonate-buffered media. Shaking incubators combine these conditions with orbital motion. Door seals, insulation and heated lids limit gradients and condensation, which otherwise produce edge effects between outer and inner wells.
What are the key features?
- Forced-air circulation for uniform temperature across stacked plates
- Humidity control from about 60 to 95 percent relative humidity
- Capacity from one plate to several dozen in racks
- Temperature range of ambient plus 5 to 60 degrees Celsius
- Stacker and robotic arm integration on larger models
- Alarms and data logging for unattended runs
What are the technical specifications?
A typical forced-air microplate incubator with humidity control configuration offers temperature range of Ambient +5 to 60 °C, temperature uniformity of ±0.5 °C, humidity range of 60–95% RH, chamber volume of 20–150 L, plate capacity of 4–60 plates (rack dependent) and recovery after door open of ≤ 10 min to 37 °C. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Forced-Air Microplate Incubator with Humidity Control is used in long ELISA incubations, immunoassay batch processing, antibody coating overnight, microbial susceptibility incubation and high-throughput plate incubation. Across the wider Microplate Incubators class, laboratories also deploy it for ELISA and immunoassay incubation, bacterial growth and susceptibility testing and mammalian cell culture in plates.
Which standards and certifications apply?
Relevant standards and methods include IEC 61010-1, IEC 61010-2-010, IEC 61326-1, ANSI/SLAS 1-2004 to 4-2004, ISO 13485 (IVD workflows) and EN 12469 (for biosafety-related enclosures, where applicable). Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install on a level surface away from drafts, direct sunlight and heat sources, with clearance for ventilation. Map temperature with calibrated probes in loaded and empty conditions at installation and periodically, and record set point drift. Use distilled or deionized water in humidity reservoirs and clean them regularly to prevent microbial growth. Clean interior surfaces with disinfectants compatible with the chamber material, and run sterilization cycles where available. Avoid frequent door openings, and allow recovery time for temperature and CO2. Calibrate CO2 sensors against a reference analyzer, and keep alarms active for unattended runs.
How do I select the right Forced-Air Microplate Incubator with Humidity Control?
Choose this variant for batches of plates needing humid, uniform incubation without CO2. Use a CO2 model for mammalian cells, as humidity alone does not maintain medium pH. In an RFQ specify temperature range, humidity range, plate capacity, recovery time and automation integration.
How does it compare with other Microplate Incubators variants?
Match the incubator to assay temperature, duration and atmosphere. Dry-block or forced-air models are enough for ELISA and binding steps at 25 to 37 degrees. Use a humidified chamber for long incubations to limit evaporation. Mammalian cell assays need CO2 and humidity control, while bacterial growth may need shaking and temperatures up to 60 degrees. Check capacity in plates, uniformity and recovery time after door opening, plus cooling below ambient for refrigerated incubations. Consider footprint, compatibility with stackers and readers, contamination control features such as HEPA filtration or copper surfaces, data logging and alarms. For regulated laboratories, verify temperature mapping documentation, calibration certificates and audit trail capabilities.
Full specifications
| Temperature range | Ambient +5 to 60 °C |
|---|---|
| Temperature uniformity | ±0.5 °C |
| Humidity range | 60–95% RH |
| Chamber volume | 20–150 L |
| Plate capacity | 4–60 plates (rack dependent) |
| Recovery after door open | ≤ 10 min to 37 °C |
| Air circulation | Forced air, adjustable fan |
| Humidity source | Water reservoir or ultrasonic mist |
| Alarm functions | Over-temperature, door, low water |
| Data interface | USB or RS-232 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Long ELISA incubations
- Immunoassay batch processing
- Antibody coating overnight
- Microbial susceptibility incubation
- High-throughput plate incubation
Key features
- Forced-air circulation for uniform temperature across stacked plates
- Humidity control from about 60 to 95 percent relative humidity
- Capacity from one plate to several dozen in racks
- Temperature range of ambient plus 5 to 60 degrees Celsius
- Stacker and robotic arm integration on larger models
- Alarms and data logging for unattended runs
Frequently asked questions about Forced-Air Microplate Incubator with Humidity Control
Humidity reduces evaporation from wells, especially at edges, which otherwise concentrates reagents and produces edge effects. Above 80 percent relative humidity, loss over overnight incubations is small for sealed plates.
Use distilled or deionized water, refill before each long run and empty and disinfect the reservoir weekly or as recommended, since stagnant water supports bacteria and fungi that can contaminate plates.
Stack height depends on rack design and airflow. Typically racks hold plates with spacing or three to five plates per stack with lids. Verify temperature mapping of the maximum load, because dense stacks slow heat equilibration.
Most ELISA steps are run at 37 degrees Celsius for 30 to 120 minutes, or at room temperature, or overnight at 2 to 8 degrees, depending on the kit. Incubators should hold the set temperature uniformly across all wells.
The edge effect is a difference in results between outer and inner wells caused by faster evaporation and temperature gradients at plate edges. Humidified chambers, sealing films and filling outer wells with liquid reduce it.
Only for mammalian cell assays in bicarbonate-buffered media, which require about 5 percent carbon dioxide to keep pH. ELISA, bacterial growth and HEPES-buffered assays do not need CO2.