LC-Ion Trap MS System
液相色谱离子阱质谱仪(LC-Ion Trap MS)
- SKU LCM-0004
- LC-MS
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- IEC 61010-1 safety (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Mass range | m/z 50–2000 (extendable to 4000) |
|---|---|
| Mass resolution | Unit; up to ~20000 in ultra-zoom |
| MS^n capability | MS^2 to MS^10 |
| Scan speed | up to 33000 Da/s (normal scan) |
| Mass accuracy | ±0.1 u (±0.2 u in MS^n) |
| Sensitivity (reserpine, 1 pg on column, full scan) | S/N ≥ 500:1 (typical) |
By Lab Manufacturers China Technical TeamLast reviewed Standards: EPA Method 1633, EPA Method 537.1, EPA Method 533
What is the LC-Ion Trap MS System?
An LC-ion trap MS stores ions in an electrodynamic trap and fragments them repeatedly, providing MSn spectra for structural elucidation. It is selected for metabolite, natural product and peptide characterization where multi-stage fragmentation matters more than high-throughput targeted quantitation.
What is the LC-Ion Trap MS System?
An LC-ion trap MS stores ions in an electrodynamic trap and fragments them repeatedly, providing MSn spectra for structural elucidation. It is selected for metabolite, natural product and peptide characterization where multi-stage fragmentation matters more than high-throughput targeted quantitation.
Overview
An LC-ion trap MS uses a three-dimensional quadrupole trap or a linear ion trap that accumulates ions from the source, isolates a chosen precursor, and fragments it by resonance-excitation collision-induced dissociation. Because fragment ions remain trapped, isolation and fragmentation can be repeated to MS^n, commonly up to MS^10, in time rather than in space. Mass range is typically m/z 50 to 2000 or 4000 with unit resolution, and ultra-zoom scans can resolve multiply charged peptide ions. Sensitivity in full scan is high because ions are accumulated, but MS/MS quantitation has lower dynamic range than triple quadrupoles because of space-charge effects. Low-mass cutoff, about one third of the precursor m/z, hides small fragments in resonance CID. The platform is compact and relatively economical for structural work.
How does it work?
In LC-MS, a high-pressure pump delivers a mobile phase through a column, typically C18, where analytes separate by their affinity for the stationary phase. The eluent, usually at 0.2 to 1 mL per minute, enters an atmospheric-pressure ion source. Electrospray ionization applies 3 to 5 kV to a nebulized stream so that charged droplets shrink and release gas-phase ions, while atmospheric-pressure chemical ionization uses a corona discharge for less polar compounds. Ions pass through a skimmer or capillary and ion guides into progressively lower-pressure regions, then into the mass analyzer. Quadrupole, ion trap, time-of-flight or hybrid analyzers then measure m/z. Mobile phases must be volatile, such as formic acid, acetic acid or ammonium formate, because non-volatile buffers foul the source.
What are the key features?
- MSn capability up to MS^10 for stepwise structural elucidation of unknown compounds
- Linear ion trap geometry with high trapping capacity and efficient ion ejection
- Automatic MS/MS triggering with dynamic exclusion for data-dependent acquisition
- Ultra-zoom scan mode to determine charge state of multiply charged ions
- Electron transfer dissociation option for peptide and post-translational modification work
- Compact design with modest vacuum and gas demands for routine laboratories
What are the technical specifications?
A typical lc-ion trap ms system configuration offers mass range of m/z 50–2000 (extendable to 4000), mass resolution of Unit; up to ~20000 in ultra-zoom, MS^n capability of MS^2 to MS^10, scan speed of up to 33000 Da/s (normal scan), mass accuracy of ±0.1 u (±0.2 u in MS^n) and sensitivity (reserpine, 1 pg on column, full scan) of S/N ≥ 500:1 (typical). Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The LC-Ion Trap MS System is used in structural elucidation of drug metabolites, natural product and plant extract characterization, peptide sequencing and post-translational modifications, glycan and lipid structural studies and teaching and research laboratories. Across the wider LC-MS class, laboratories also deploy it for drug impurity profiling and metabolite identification, bioanalysis of drugs in plasma and urine and pesticide and veterinary drug residue testing.
Which standards and certifications apply?
Relevant standards and methods include EPA Method 1633, EPA Method 537.1, EPA Method 533, USP <621> Chromatography, FDA Bioanalytical Method Validation Guidance, ICH M10 Bioanalytical Method Validation, EN 15662, CLSI C62-A, ISO/IEC 17025 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and IEC 61010-1 safety; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Provide a nitrogen supply of at least 99.5 percent purity, either from a generator delivering about 20 to 35 L per minute or from liquid supply, plus a fume extraction line for source exhaust. Use LC-MS grade solvents and additives, since impurities raise background and suppress ionization. Calibrate the mass scale regularly with the manufacturer's calibrant solution, and run system suitability injections before each batch. Clean the spray shield, capillary and ion transfer optics when sensitivity drops, and replace rotary pump oil on schedule. Monitor for ion suppression through post-column infusion during method development. Regulated laboratories need software with audit trails and electronic signatures compatible with 21 CFR Part 11, and instrument qualification to IQ, OQ and PQ protocols.
How do I select the right LC-Ion Trap MS System?
Select an LC-ion trap MS for structural elucidation requiring multi-stage fragmentation at moderate cost. Choose Q-TOF if accurate mass is essential, or triple quadrupole for regulated quantitation. In an RFQ, specify MS^n depth, scan speed, mass range, fragmentation options such as ETD, and helium supply requirements.
How does it compare with other LC-MS variants?
The decisive choice is the mass analyzer. Single quadrupole LC-MS confirms molecular weight and performs simple quantitation at a modest cost. Triple quadrupole instruments deliver the best sensitivity for targeted quantitation in bioanalysis, residues and clinical assays. Q-TOF systems provide accurate mass and MS/MS of unknowns, suited to impurity profiling, metabolite identification and screening. Ion traps offer multi-stage fragmentation for structural work. Mass-directed purification systems add fraction collection. Then check the ion source options such as ESI, APCI and APPI, scan speed for UHPLC peaks below 3 seconds wide, polarity switching speed, and the robustness of the source with dirty matrices. Confirm compatibility with your LC pumps, available software for regulated work, and nitrogen consumption for sheath and drying gases.
Full specifications
| Mass range | m/z 50–2000 (extendable to 4000) |
|---|---|
| Mass resolution | Unit; up to ~20000 in ultra-zoom |
| MS^n capability | MS^2 to MS^10 |
| Scan speed | up to 33000 Da/s (normal scan) |
| Mass accuracy | ±0.1 u (±0.2 u in MS^n) |
| Sensitivity (reserpine, 1 pg on column, full scan) | S/N ≥ 500:1 (typical) |
| Isolation width | 0.3–10 u |
| Fragmentation modes | Resonance CID; optional ETD |
| Ion sources | ESI, APCI (nanospray optional) |
| Trap gas | Helium, 99.999% |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Structural elucidation of drug metabolites
- Natural product and plant extract characterization
- Peptide sequencing and post-translational modifications
- Glycan and lipid structural studies
- Teaching and research laboratories
Key features
- MSn capability up to MS^10 for stepwise structural elucidation of unknown compounds
- Linear ion trap geometry with high trapping capacity and efficient ion ejection
- Automatic MS/MS triggering with dynamic exclusion for data-dependent acquisition
- Ultra-zoom scan mode to determine charge state of multiply charged ions
- Electron transfer dissociation option for peptide and post-translational modification work
- Compact design with modest vacuum and gas demands for routine laboratories
Frequently asked questions about LC-Ion Trap MS System
MSn means successive stages of mass spectrometry. An ion is isolated and fragmented in MS^2, a product ion is isolated and fragmented again in MS^3, and so on. Each stage reveals more of the structure, which helps map fragmentation pathways of unknowns.
Trapped ions repel each other, so space-charge effects limit the number of ions stored and reduce the linear dynamic range. Scan-based MS/MS also has lower duty cycle for many compounds than MRM on a triple quadrupole. Ion traps are therefore mainly used for qualitative work.
In resonance-excitation CID, fragment ions with m/z below about 28 to 30 percent of the precursor m/z are not stably trapped and are lost. This hides low-mass diagnostic fragments, which some instruments mitigate with pulsed-Q or higher-energy collision dissociation.
Electrospray ionization is the most common and suits polar and ionizable compounds from small molecules to proteins. APCI suits less polar, thermally stable analytes below about 1500 Da. APPI handles nonpolar compounds such as PAHs. Many sources allow switching or dual-mode operation.
Ion suppression is a reduction in analyte signal caused by co-eluting matrix components that compete for charge or disrupt droplet formation. It affects accuracy and sensitivity. It is managed with better sample cleanup, chromatographic separation, dilution and isotopically labeled internal standards.
Use volatile additives such as formic acid, acetic acid, ammonium formate, ammonium acetate and ammonium hydroxide, generally at 0.1 percent or 2 to 10 mM. Avoid non-volatile phosphate, borate and high concentrations of ion-pairing reagents, which contaminate the source and suppress ionization.