Automated ELISA Processing System
全自动酶免分析系统
- SKU ELS-0004
- ELISA Systems
- CE (to be confirmed)
- ISO 13485 manufacturing (to be confirmed)
- NMPA class II registration (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Throughput | Up to 400–1200 wells/h depending on protocol |
|---|---|
| Incubation chambers | 2–4 chambers, 25–50 °C ±0.5 °C |
| Sample volume | 10–50 µL per well |
| Sample positions | 60–180 tube positions with barcode |
| Reagent positions | Up to 40 cooled positions at 4–10 °C |
| Plate capacity | 10–40 plates onboard |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 15189, CLSI EP05, CLSI I/LA20
What is the Automated ELISA Processing System?
An automated ELISA processing system performs sample and reagent pipetting, incubation, washing, substrate addition and reading without operator intervention. It suits hospital and reference laboratories processing several hundred to several thousand wells per day with many different kit protocols.
What is the Automated ELISA Processing System?
An automated ELISA processing system performs sample and reagent pipetting, incubation, washing, substrate addition and reading without operator intervention. It suits hospital and reference laboratories processing several hundred to several thousand wells per day with many different kit protocols.
Overview
The system integrates a pipetting robot, one or more incubation chambers, a plate washer and a reader, all controlled by software that holds the assay protocols. Plates are moved by a robotic arm or track, samples are transferred from tubes into plates with disposable tips or a fixed probe, and reagents are added from cooled positions. Walkaway time depends on the number of incubation steps and can reach several hours, but the sample volume of serum or plasma needed is small, typically 10 to 50 microliters per well. Traceability features include positive sample identification, reagent lot capture, control position verification and an audit trail. Because several kits may run in parallel, the software validates protocol compatibility, temperature and timing before starting a run, and it reports results as concentrations, cut-off indices or qualitative interpretations.
How does it work?
Enzyme-linked immunosorbent assay uses a plastic microplate whose wells carry a bound capture antibody or antigen. Sample is added, and any target present binds to the solid phase during incubation. After washing, a detection antibody conjugated to an enzyme such as horseradish peroxidase or alkaline phosphatase is added and binds to the captured target. A further wash removes unbound conjugate, then a substrate is added and the enzyme converts it into a coloured product. The reaction is stopped with acid or base and the absorbance is read at a defined wavelength, usually 450 nanometers with a reference wavelength between 600 and 650 nanometers. Absorbance is proportional to the amount of bound analyte, and a calibration curve, often four-parameter logistic, converts readings into concentrations or index values. Strict washing, incubation timing and temperature control determine the reliability of the result.
What are the key features?
- Robotic pipetting with disposable tips or fixed probes
- Integrated incubation chambers, washer and reader
- Walkaway operation of several hours for long protocols
- Positive sample identification with reagent lot traceability
- Protocol compatibility checks before a run starts
- Audit trail with operator and reagent data
What are the technical specifications?
A typical automated elisa processing system configuration offers throughput of Up to 400–1200 wells/h depending on protocol, incubation chambers of 2–4 chambers, 25–50 °C ±0.5 °C, sample volume of 10–50 µL per well, sample positions of 60–180 tube positions with barcode, reagent positions of Up to 40 cooled positions at 4–10 °C and plate capacity of 10–40 plates onboard. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Automated ELISA Processing System is used in hospital infectious disease serology at volume, blood donor screening laboratories, autoimmune and allergy panels and reference laboratories running many kit protocols. Across the wider ELISA Systems class, laboratories also deploy it for infectious disease serology, allergy specific IgE testing and autoimmune antibody screening.
Which standards and certifications apply?
Relevant standards and methods include ISO 15189, CLSI EP05, CLSI I/LA20, CLSI M29, IEC 61010-1 and IEC 61010-2-101. Certification candidates for this model are CE, ISO 13485 manufacturing and NMPA class II registration; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the reader and washer on a level bench away from vibration and direct light, and keep the plate carrier and optical window free of dust. Verify performance at installation with a reference dye plate for optical accuracy and with controls for each assay. Daily tasks include checking the washer manifold for blockage, priming the wash lines and running controls at two or three levels. Plates should be read immediately after stopping the reaction, since colour continues to develop and delays bias results, and consistent timing across the plate matters more than absolute speed. Replace tubing and manifold pins when flow becomes uneven, and dispose of wash waste and plates as biohazardous material.
How do I select the right Automated ELISA Processing System?
Specify an automated processor when daily volume exceeds about five hundred wells, when many protocols run in parallel, or when traceability and walkaway operation are required. In the RFQ state the number of kits and their incubation requirements, the plate throughput for your most demanding protocol, sample and reagent position counts, and whether the software can manage kit lot changes without reprogramming.
How does it compare with other ELISA Systems variants?
Define the assays first, then decide how much automation is appropriate. Laboratories running fewer than one hundred wells per day can work with a reader and washer plus manual pipetting, while larger volumes justify an automated processor with plate handling, incubation and reporting. Compare readers by wavelength range, filter set, linearity, and whether they support kinetic and dual-wavelength measurement. For washers, check residual volume, the number of wash cycles programmable, and the manifolds for 96-well and strip formats. For automated systems, check plate capacity, walkaway time, whether the software handles assay protocols centrally, and the traceability of each step with lot numbers and operator identification. Confirm the assay registration status in your market, since ELISA kits are regulated separately from the instruments.
Full specifications
| Throughput | Up to 400–1200 wells/h depending on protocol |
|---|---|
| Incubation chambers | 2–4 chambers, 25–50 °C ±0.5 °C |
| Sample volume | 10–50 µL per well |
| Sample positions | 60–180 tube positions with barcode |
| Reagent positions | Up to 40 cooled positions at 4–10 °C |
| Plate capacity | 10–40 plates onboard |
| Residual volume | < 3 µL per well after washing |
| Walkaway time | 2–8 h depending on protocol |
| Interfaces | LAN with HL7 or ASTM two-way communication |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Hospital infectious disease serology at volume
- Blood donor screening laboratories
- Autoimmune and allergy panels
- Reference laboratories running many kit protocols
Key features
- Robotic pipetting with disposable tips or fixed probes
- Integrated incubation chambers, washer and reader
- Walkaway operation of several hours for long protocols
- Positive sample identification with reagent lot traceability
- Protocol compatibility checks before a run starts
- Audit trail with operator and reagent data
Frequently asked questions about Automated ELISA Processing System
The software stores each kit protocol as a program with defined pipetting, incubation and washing steps, and checks compatibility before a run. Laboratories must validate each new protocol and kit lot, including control positions and cut-off calculation, before reporting patient results.
Fixed probes reduce consumable cost and are common on integrated platforms, but they need effective washing and carryover verification between samples. Disposable tips simplify carryover control at higher cost per test, and some instruments allow both modes depending on the assay.
The software records the lot number and expiry of every reagent position and usually requires calibration or control verification when a new lot is loaded. The laboratory should define acceptance criteria for each lot change and keep the control results with the run record for audit purposes.
The single point of failure: if the robot or one module stops, all protocols on that platform stop. Mitigation includes a service contract with a defined response time, a standby manual reader and washer, and a written downtime procedure that identifies which tests are urgent.
A reader only measures absorbance after the assay has been performed manually, while an automated system also handles sample and reagent pipetting, incubation, washing, substrate addition and timing. Readers suit low volumes and flexible menus; automated systems reduce variability and labour at higher volumes.
Reading at the assay wavelength and subtracting a reference wavelength between 600 and 650 nanometers corrects for scratches, dust and bubbles on the plate. It reduces well-to-well variation and is standard practice for quantitative ELISA, especially at low absorbance values where optical artefacts have more influence.
Check the manifold for blocked or misaligned pins before each run, and verify residual volume weekly with a gravimetric or dye test. Uneven aspiration leaves wash solution that carries over unbound conjugate, which raises background and reduces precision, so any change in background should prompt a washer check.