Benchtop Conductivity Meter
台式电导率仪
- SKU COM-0001
- Conductivity Meters
- CE (to be confirmed)
- RoHS (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Conductivity range | 0.001 µS/cm–1000 mS/cm |
|---|---|
| Accuracy | ±0.5% of reading |
| Resolution | 0.001 µS/cm (lowest range) |
| Cell constants | 0.01–10 cm⁻¹ |
| Resistivity range | up to 20 MΩ·cm |
| Salinity range | 0–80 PSU |
By Lab Manufacturers China Technical TeamLast reviewed Standards: USP <645>, Ph. Eur. 2.2.38, ASTM D1125
What is the Benchtop Conductivity Meter?
A benchtop conductivity meter is a mains-powered laboratory instrument with automatic ranging, selectable cell constants and calibration with traceable standards. It gives accuracy of ±0.5 percent of reading and is used for water purity, QC and research where recorded, repeatable results are needed.
What is the Benchtop Conductivity Meter?
A benchtop conductivity meter is a mains-powered laboratory instrument with automatic ranging, selectable cell constants and calibration with traceable standards. It gives accuracy of ±0.5 percent of reading and is used for water purity, QC and research where recorded, repeatable results are needed.
Overview
The benchtop meter covers a wide conductivity span, typically 0.001 microsiemens per centimeter to 1000 millisiemens per centimeter, by automatic range selection and support for cell constants from 0.01 to 10 per centimeter. It offers linear, nonlinear and natural-water temperature compensation, and can display conductivity, resistivity, salinity and TDS. Calibration can be performed with one to several points and the instrument recognizes standard solutions automatically. Accuracy of ±0.5 percent of reading is typical with a four-electrode or two-electrode graphite or platinum cell. GLP functions include calibration logs, cell constant storage, user identification and data export by USB or LIMS. A stirrer, electrode stand and optional second channel for pH are common accessories. Laboratories use it in pharmaceutical QC, environmental testing, food analysis and education.
How does it work?
Conductivity is measured by applying an alternating voltage across two or four electrodes of a cell with known geometry and measuring the resulting current. The conductance, in siemens, multiplied by the cell constant (the ratio of electrode spacing to area, in reciprocal centimeters) gives conductivity in microsiemens or millisiemens per centimeter. Alternating current at frequencies from hundreds of hertz to several kilohertz avoids electrode polarization. Conductivity rises by about 2 percent per degree Celsius, so meters compensate to a reference temperature, usually 25 degrees Celsius, using a linear coefficient or a nonlinear function for natural or ultrapure water. Cells with constants of 0.01 to 0.1 per centimeter are used for pure water, 1 per centimeter for general work and 10 per centimeter for concentrated solutions. Inductive cells use two toroidal coils and have no exposed electrodes, which suits aggressive or fouling media. TDS is estimated by multiplying conductivity by a factor of about 0.5 to 0.7.
What are the key features?
- Automatic range selection from 0.001 µS/cm to 1000 mS/cm
- Support for cell constants from 0.01 to 10 cm⁻¹ with stored cell data
- Linear, nonlinear and natural-water temperature compensation options
- Display of conductivity, resistivity, salinity and TDS on one screen
- Automatic standard recognition with single- or multi-point calibration
- GLP logging and USB or LIMS data export
What are the technical specifications?
A typical benchtop conductivity meter configuration offers conductivity range of 0.001 µS/cm–1000 mS/cm, accuracy of ±0.5% of reading, resolution of 0.001 µS/cm (lowest range), cell constants of 0.01–10 cm⁻¹, resistivity range of up to 20 MΩ·cm and salinity range of 0–80 PSU. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Benchtop Conductivity Meter is used in pharmaceutical water QC, environmental water analysis, food and beverage testing, research and teaching labs and cleaning validation rinse water. Across the wider Conductivity Meters class, laboratories also deploy it for pharmaceutical purified water testing, drinking and wastewater monitoring and boiler and cooling water control.
Which standards and certifications apply?
Relevant standards and methods include USP <645>, Ph. Eur. 2.2.38, ASTM D1125, ASTM D5391, ISO 7888, OIML R 68, IEC 60746-3 and GB/T 6908. Certification candidates for this model are CE, RoHS and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Calibrate with traceable potassium chloride standards close to the sample range, for example 84, 1413 or 12880 microsiemens per centimeter, at a controlled temperature. For ultrapure water, measure in a flow-through cell to avoid absorption of carbon dioxide, which raises conductivity within minutes. Rinse the cell with sample before measuring and store it dry or in deionized water as recommended. Platinum black coating should be protected from contamination and re-platinized when needed. Verify the cell constant periodically. Pharmaceutical laboratories should follow USP <645> or Ph. Eur. 2.2.38 staged testing and keep calibration and verification records. Keep cables away from power lines to avoid interference and ensure samples are not contaminated by handling.
How do I select the right Benchtop Conductivity Meter?
Choose a benchtop meter for accurate, documented laboratory work and pharmacopoeial testing. It is not needed for rough field checks. In an RFQ specify conductivity range, required accuracy, cell constants, temperature compensation modes, GLP and data output needs and any pH or second-channel option.
How does it compare with other Conductivity Meters variants?
Start with the conductivity range of your samples, because the cell constant must match it: use 0.01 to 0.1 per centimeter for ultrapure water, 1 per centimeter for drinking water and most laboratory work, and 10 per centimeter for brines and concentrated solutions. Then choose the format. Benchtop meters give the best accuracy and GLP features, portable meters suit field sampling and plant rounds, pen testers suit screening, and online analyzers give continuous monitoring with relay and 4–20 mA outputs. Pharmaceutical water testing requires a meter and cell that support USP <645> with temperature compensation disabled and calibration traceable to standards. For fouling or corrosive media consider inductive sensors. Check the accuracy specification, temperature compensation options, cell material and the cost of calibration standards.
Full specifications
| Conductivity range | 0.001 µS/cm–1000 mS/cm |
|---|---|
| Accuracy | ±0.5% of reading |
| Resolution | 0.001 µS/cm (lowest range) |
| Cell constants | 0.01–10 cm⁻¹ |
| Resistivity range | up to 20 MΩ·cm |
| Salinity range | 0–80 PSU |
| TDS range | 0–500 g/L |
| Temperature range | −5 to 105 °C |
| Reference temperature | 20 or 25 °C selectable |
| Data storage | ≥ 1000 results |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Pharmaceutical water QC
- Environmental water analysis
- Food and beverage testing
- Research and teaching labs
- Cleaning validation rinse water
Key features
- Automatic range selection from 0.001 µS/cm to 1000 mS/cm
- Support for cell constants from 0.01 to 10 cm⁻¹ with stored cell data
- Linear, nonlinear and natural-water temperature compensation options
- Display of conductivity, resistivity, salinity and TDS on one screen
- Automatic standard recognition with single- or multi-point calibration
- GLP logging and USB or LIMS data export
Frequently asked questions about Benchtop Conductivity Meter
Typical accuracy is ±0.5 percent of reading for the electronics with a well-characterized cell, though the overall uncertainty also depends on cell constant determination, standard solution accuracy and temperature. Calibrate in the range of your samples to minimize error.
Many benchtop meters support USP <645> by allowing uncompensated conductivity readings, a cell constant verified within 2 percent and staged testing of water. Compliance must be confirmed for the specific model, cell and validation procedure used by the laboratory.
Yes, dual-channel benchtop models accept both a conductivity cell and a pH electrode. Measure them in separate aliquots for ultrapure water because electrodes and dissolved carbon dioxide can change conductivity within minutes.
It measures how well a solution conducts electricity, reported in microsiemens or millisiemens per centimeter. Because dissolved ions carry the current, conductivity indicates total ionic content and is used for water purity, salinity and TDS estimation.
TDS meters measure conductivity and multiply it by a conversion factor, commonly 0.5 to 0.7, to estimate dissolved solids in mg/L. The factor depends on the ion mix, so the result is an estimate unless verified against gravimetric TDS.
Choose a cell constant that places the sample in the middle of the cell range: 0.01 to 0.1 per centimeter for ultrapure water below 20 microsiemens per centimeter, 1 per centimeter for 10 to 20,000 microsiemens per centimeter, and 10 per centimeter for higher values.