Class II Type B2 Biological Safety Cabinet
II 级 B2 型生物安全柜
- SKU CCH-0002
- Cell Culture Hoods
- CE (to be confirmed)
- NSF/ANSI 49 certification (to be confirmed)
- EN 12469 certification (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Working width | 1.2–1.8 m |
|---|---|
| Exhaust | 100% hard-ducted to external blower |
| Downflow velocity | approx. 0.25–0.35 m/s |
| Inflow velocity | approx. 0.5 m/s |
| Filtration | HEPA, 99.99% at 0.3 µm, supply and exhaust |
| Duct static pressure | as specified, typically 250–500 Pa |
By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 12469, NSF/ANSI 49, ISO 14644-1
What is the Class II Type B2 Biological Safety Cabinet?
A class II type B2 biosafety cabinet passes all of its air through HEPA filters and exhausts it outdoors, so no air is recirculated. It is used where volatile chemicals, radionuclides or hazardous aerosols must be fully removed.
What is the Class II Type B2 Biological Safety Cabinet?
A class II type B2 biosafety cabinet passes all of its air through HEPA filters and exhausts it outdoors, so no air is recirculated. It is used where volatile chemicals, radionuclides or hazardous aerosols must be fully removed.
Overview
The cabinet draws room air and its own filtered supply air through the work zone and discharges everything through an exhaust HEPA filter to an external duct, maintaining a hard-ducted 100% exhaust configuration. Because nothing recirculates, chemical vapours and radioactive aerosols cannot accumulate in the plenums or filters, which makes the type suitable for work with volatile solvents, tracer radionuclides and cytotoxic drugs handled alongside cells. Airflow balance is more critical than in type A2 cabinets, and installation requires a dedicated exhaust system with specified static pressure and a monitoring device. Downflow and inflow velocities are similar to type A2, and certification procedures follow EN 12469 or NSF/ANSI 49 as for other class II cabinets.
How does it work?
Sterile work depends on supplying particle-free air and preventing entry of contaminants from the room. Room air is drawn through a HEPA filter that removes 99.99% of particles at 0.3 µm, and the filtered air flows in a laminar stream over the work surface. In a clean bench, air moves horizontally or vertically across the surface and out toward the operator, protecting the sample only. In a class II biological safety cabinet, air enters a front grille, passes through the supply filter to create a sterile downflow curtain, and is then split so that part recirculates and part is exhausted through a second HEPA filter, protecting both product and operator. A front air barrier prevents room air from entering the work zone when arms reach in, and airflow velocities are designed to balance containment against the risk of disturbing cultures.
What are the key features?
- 100% exhaust through HEPA filters with no recirculation
- Hard-ducted exhaust to an external blower system
- Suitable for volatile solvents and tracer radionuclides
- Downflow and inflow velocities matched to class II requirements
- Exhaust airflow monitoring with alarm
- Stainless steel interior with decontamination access points
What are the technical specifications?
A typical class ii type b2 biological safety cabinet configuration offers working width of 1.2–1.8 m, exhaust of 100% hard-ducted to external blower, downflow velocity of approx. 0.25–0.35 m/s, inflow velocity of approx. 0.5 m/s, filtration of HEPA, 99.99% at 0.3 µm, supply and exhaust and duct static pressure of as specified, typically 250–500 Pa. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Class II Type B2 Biological Safety Cabinet is used in cell work with volatile solvents or fixatives, cytotoxic drug handling with cell cultures, radionuclide tracer experiments in cells and cell processing under full containment. Across the wider Cell Culture Hoods class, laboratories also deploy it for aseptic cell culture and subculture, primary cell and stem cell handling and media preparation and bottle filling.
Which standards and certifications apply?
Relevant standards and methods include EN 12469, NSF/ANSI 49, ISO 14644-1, IEC 61010-1, IEC 61326-1 and USP <1046>. Certification candidates for this model are CE, NSF/ANSI 49 certification, EN 12469 certification and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
A new or relocated cabinet must be certified after installation and yearly thereafter, with filter integrity, downflow velocity, inflow velocity and airflow visualisation recorded against EN 12469 or NSF/ANSI 49. Avoid placing the cabinet near doors, windows, air supply diffusers or high-traffic routes, because draughts disturb the air barrier. Clean the work surface before and after use with 70% ethanol or a compatible disinfectant, and run the blower for several minutes before starting work. Do not block the front or rear grilles with equipment, since that alters airflow and can break containment. UV lamps are a supplement, not a substitute, for cleaning, and should be replaced when intensity falls below specification. Record certification, cleaning and maintenance in a log for audit purposes.
How do I select the right Class II Type B2 Biological Safety Cabinet?
Select a type B2 cabinet only when volatile chemicals, radionuclides or hazardous aerosols must be removed completely, since it requires an external exhaust system. Specify exhaust static pressure, blower capacity, duct routing, monitoring and alarm features and the containment level required.
How does it compare with other Cell Culture Hoods variants?
Decide first whether the operator needs protection. Any work with human or animal cells, infectious material or genetically modified organisms requires a class II cabinet, and a type A2 model covers most cell culture needs when recirculating exhaust is acceptable. Type B2 cabinets exhaust all air to the outside and are used with volatile chemicals or radionuclides that must not recirculate. Clean benches are appropriate only for non-hazardous cultures where sample protection alone is sufficient. Then check working dimensions, airflow velocities and alarms, filter integrity test points, interior material, UV lamp, and whether an electrical outlet panel and gas or vacuum service are needed inside the work zone. Confirm that the cabinet will be certified after installation in your laboratory and that the annual certification service is available locally.
Full specifications
| Working width | 1.2–1.8 m |
|---|---|
| Exhaust | 100% hard-ducted to external blower |
| Downflow velocity | approx. 0.25–0.35 m/s |
| Inflow velocity | approx. 0.5 m/s |
| Filtration | HEPA, 99.99% at 0.3 µm, supply and exhaust |
| Duct static pressure | as specified, typically 250–500 Pa |
| Exhaust airflow monitor | Continuous with alarm |
| Volatile chemical compatibility | Yes, with no recirculation |
| Noise level | below 67 dB(A) |
| Dimensions (W x D x H) | approx. 1,400 x 800 x 1,500 mm |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Cell work with volatile solvents or fixatives
- Cytotoxic drug handling with cell cultures
- Radionuclide tracer experiments in cells
- Cell processing under full containment
Key features
- 100% exhaust through HEPA filters with no recirculation
- Hard-ducted exhaust to an external blower system
- Suitable for volatile solvents and tracer radionuclides
- Downflow and inflow velocities matched to class II requirements
- Exhaust airflow monitoring with alarm
- Stainless steel interior with decontamination access points
Frequently asked questions about Class II Type B2 Biological Safety Cabinet
Because all the air leaves through the exhaust HEPA filter, it must go to an external blower and discharge point. There is no recirculation path, so chemical vapours cannot concentrate in the cabinet's plenums or filters.
A dedicated exhaust system sized for the cabinet's airflow and static pressure, a duct route with adequate length and few bends, an accessible blower, and airflow monitoring with alarm. Ductwork and blower sizing should be confirmed by the supplier before installation.
Yes. Because exhaust is driven by an external blower, changes in duct pressure affect cabinet airflow, so a monitoring device and periodic verification are essential. Building ventilation changes should be assessed for their effect on the cabinet.
Use a class II biosafety cabinet whenever cultures may contain human or animal cells, infectious agents or genetically modified material, because it protects the operator as well as the sample. A clean bench is acceptable only for non-hazardous cultures needing sample protection.
At installation, after any relocation or major repair, and at least annually thereafter. Certification covers HEPA filter integrity, downflow and inflow velocities, and airflow visualisation against EN 12469 or NSF/ANSI 49, with results recorded.
Away from doors, windows, air conditioning diffusers and busy walkways, because cross-draughts disturb the air barrier and can draw contaminants into the work zone. Confirm suitable bench height and clearance above the cabinet for exhaust and service access.