High-Vacuum Volumetric Physisorption Analyzer for Micropore and Mesopore Analysis
微孔与介孔分析用高真空容量法物理吸附分析仪
- CE (to be confirmed)
- +2
Welcome to Lab Manufacturers China — laboratory instruments direct from a Chinese manufacturer

薄膜与平整基底用氪气低比表面积分析仪
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
| Adsorbate | Krypton at 77 K |
|---|---|
| Minimum total surface area | ≥ 0.05 m² |
| Surface area range | 0.001–20 m²/g |
| Transducer range | 1 torr and 10 torr |
| Transducer accuracy | ±0.15% of reading |
| Relative pressure for BET | 0.05–0.30 |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 9277, ISO 15901-2, ISO 15901-3
A volumetric analyzer configured for krypton adsorption at 77 K with a low-volume manifold and a sensitive low-range transducer. It measures total areas as low as 0.05 m², so it suits thin films, sintered ceramics, fuel pellets, coated substrates and other low-area samples.
A volumetric analyzer configured for krypton adsorption at 77 K with a low-volume manifold and a sensitive low-range transducer. It measures total areas as low as 0.05 m², so it suits thin films, sintered ceramics, fuel pellets, coated substrates and other low-area samples.
With nitrogen, the signal from samples below 1 m² of total area is too small to separate from the gas in the free space. Krypton has a saturation pressure of only about 2.5 torr at 77 K, so the same number of adsorbed molecules produces a far larger relative change in pressure. This variant minimizes manifold volume, uses a pressure transducer of 1 or 10 torr full scale with high resolution, and applies corrections for the solid-phase saturation pressure of krypton. The BET analysis is done at 0.05 to 0.30 relative pressure, following ISO 9277 and ASTM D5604 style practice. Sample tubes are designed with filler rods to reduce dead volume. Typical repeatability is within 3 to 5 percent at low areas, larger than for nitrogen on high-area powders.
These analyzers expose a degassed solid to an adsorbate and measure how much is taken up. In static volumetric physisorption, known doses of nitrogen, argon or krypton are admitted to a manifold and sample tube at 77 K or 87 K, and the pressure drop after equilibration gives the adsorbed volume. Isotherms are analyzed with BET for surface area, t-plot and DFT methods for micropores, and BJH for mesopores. Gravimetric systems weigh the sample directly while vapor pressure is controlled. Chemisorption analyzers dose hydrogen, CO or oxygen at elevated temperature and use thermal conductivity detection for temperature-programmed reduction and desorption. Key components are vacuum pumps, pressure transducers, a cryogenic bath or furnace, and analysis software.
A typical krypton low-surface-area analyzer for thin films and flat substrates configuration offers adsorbate of Krypton at 77 K, minimum total surface area of ≥ 0.05 m², surface area range of 0.001–20 m²/g, transducer range of 1 torr and 10 torr, transducer accuracy of ±0.15% of reading and relative pressure for BET of 0.05–0.30. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
The Krypton Low-Surface-Area Analyzer for Thin Films and Flat Substrates is used in thin film and coating surface area, nuclear fuel pellet and ceramic analysis, flat wafer and substrate characterization and low-area pharmaceutical crystals. Across the wider Surface Area Analyzers class, laboratories also deploy it for catalyst and support characterization, battery electrode powder analysis and activated carbon and adsorbent testing.
Relevant standards and methods include ISO 9277, ISO 15901-2, ISO 15901-3, ASTM D3663 and ASTM D4365. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Sample degassing is the main source of error; select temperature and time so that moisture and volatiles are removed without changing the structure, and weigh the sample after degassing. Use adequate sample mass to provide at least 1 m² of total surface area, or 0.05 m² with krypton. Check free-space calibration and leak rates regularly. Ensure a reliable supply of high-purity gases, typically 99.999 percent, and monitor liquid nitrogen levels. Verify performance with certified reference materials of known surface area. Place the unit on a stable bench with ventilation for cryogenic and hydrogen work, and follow local rules for compressed gas cylinders.
Select this variant when samples have less than about 1 m² total area, or when the specification is written in cm² or m² for a whole article rather than per gram. A standard nitrogen analyzer with a krypton option may be adequate for occasional use. In an RFQ, specify sample mass range, krypton purity, sample tube size and whether reporting follows ISO 9277.
Decide first what must be measured. Surface area and mesopore size of powders, catalysts and electrode materials suit a volumetric physisorption analyzer; micropore work requires turbomolecular pumping and low-pressure transducers. Very low area films and substrates need krypton capability. Water or solvent uptake calls for a gravimetric vapor sorption analyzer. Metal dispersion and acid site measurements need a chemisorption system with a furnace and detector. Hydrogen and methane storage research needs a high-pressure sorption analyzer. Compare the number of analysis ports, degassing stations, pressure range, transducer accuracy, liquid nitrogen consumption and compliance with ISO 9277 and ISO 15901-2 when specifying.
| Adsorbate | Krypton at 77 K |
|---|---|
| Minimum total surface area | ≥ 0.05 m² |
| Surface area range | 0.001–20 m²/g |
| Transducer range | 1 torr and 10 torr |
| Transducer accuracy | ±0.15% of reading |
| Relative pressure for BET | 0.05–0.30 |
| Repeatability | ≤ 4% RSD at 0.1 m² total |
| Analysis ports | 1–2 |
| Degassing temperature | Ambient to 350 °C |
| Vacuum system | Rotary vane + turbomolecular pump |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Krypton has a very low saturation pressure at 77 K, about 2.5 torr, so a small number of adsorbed molecules changes the relative pressure measurably. Nitrogen would adsorb too little to be distinguished from gas in the free space.
They agree reasonably on many non-porous solids but can differ by several percent because the krypton molecular cross-sectional area is less certain, commonly taken as 0.21 nm². Report the adsorbate and cross-section used with every result.
The limit is set by free-space volume, transducer resolution and temperature stability. With a low-volume krypton manifold, total areas of about 0.05 m² can be measured with acceptable repeatability, below which uncertainty rises sharply.
A BET analyzer measures gas adsorption on a solid at cryogenic temperature and applies the Brunauer–Emmett–Teller model to calculate specific surface area in m²/g. Most instruments also give pore volume and pore size distribution from the same isotherm.
Degassing removes water, solvents and adsorbed gases that would otherwise occupy the surface and bias the measured area. Temperature and time depend on the material, for example 150–300 °C for many oxides, and must not collapse the pore structure.
Not usually. Physisorption analyzers measure area and pores using nitrogen at 77 K, while metal dispersion requires chemisorption of hydrogen or CO at controlled temperatures with a furnace and detector. Some combined platforms offer both modules in one cabinet.