Micro Carbon Residue Tester for Fuel Oil
燃料油微量残炭测定仪
- SKU FOA-0004
- Fuel Oil Analysis
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Residue range | 0.01–30 % m/m |
|---|---|
| Final temperature | 500 °C |
| Temperature accuracy | ± 2 °C |
| Nitrogen flow | 600 ± 50 mL/min |
| Number of positions | 6 or 12 |
| Sample mass | ≈ 0.15–5 g (by residue level) |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 8217, ASTM D396, ASTM D4052
What is the Micro Carbon Residue Tester for Fuel Oil?
A micro carbon residue tester thermally cracks a weighed fuel sample under nitrogen at 500 degrees Celsius and weighs the remaining carbonaceous residue. It predicts deposit-forming tendency in engines and boilers and is required by ISO 8217 for residual marine fuels.
What is the Micro Carbon Residue Tester for Fuel Oil?
A micro carbon residue tester thermally cracks a weighed fuel sample under nitrogen at 500 degrees Celsius and weighs the remaining carbonaceous residue. It predicts deposit-forming tendency in engines and boilers and is required by ISO 8217 for residual marine fuels.
Overview
About 1 g of fuel is weighed into a glass vial and placed in a heated aluminum block, which is ramped to 500 degrees Celsius under a nitrogen flow of 600 mL/min. During a 30-minute cycle, volatile fractions evaporate or crack, and the nonvolatile residue coke remains. After cooling in nitrogen, vials are weighed and the residue is expressed as mass percent. The method is an alternative to Conradson and Ramsbottom carbon residue tests, correlates closely with them, and requires a smaller sample, less time and less operator attention. Modern units heat 6 to 12 samples simultaneously in a single run, control nitrogen accurately and automatically adjust the thermal profile according to ASTM D4530 or ISO 10370.
How does it work?
Fuel oil specifications rely on a set of physical and chemical tests. Density is obtained by oscillating U-tube: the natural frequency of a tube filled with oil depends on its mass, so period is converted to density at 15 degrees Celsius. Sulfur is measured by energy-dispersive X-ray fluorescence, where X-rays excite sulfur atoms and the intensity of the S K-alpha line, calibrated against standards, gives mass percent. Flash point is determined by heating oil in a closed Pensky-Martens cup with stirring and applying an ignition source at intervals until the vapor flashes. Micro carbon residue is obtained by thermally decomposing a weighed sample under nitrogen at 500 degrees Celsius. Water and sediment are separated by centrifuging fuel with solvent and reading the volume of the settled layer.
What are the key features?
- Heating block with precise 500 °C profile and ramp control
- Nitrogen flow control at 600 mL/min with mass flow controller
- Simultaneous testing of 6 to 12 sample vials
- Automatic cooling cycle under nitrogen atmosphere
- Built-in ASTM D4530 and ISO 10370 test programs
What are the technical specifications?
A typical micro carbon residue tester for fuel oil configuration offers residue range of 0.01–30 % m/m, final temperature of 500 °C, temperature accuracy of ± 2 °C, nitrogen flow of 600 ± 50 mL/min, number of positions of 6 or 12 and sample mass of ≈ 0.15–5 g (by residue level). Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Micro Carbon Residue Tester for Fuel Oil is used in marine residual fuel qualification, crude oil and residue quality assessment and diesel fuel carbon residue on 10 % bottoms. Across the wider Fuel Oil Analysis class, laboratories also deploy it for bunker fuel quality verification, power plant fuel inspection and refinery blending control.
Which standards and certifications apply?
Relevant standards and methods include ISO 8217, ASTM D396, ASTM D4052, ASTM D4294, ISO 8754, ASTM D93, ISO 2719, ASTM D4530 and ASTM D4007. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Fuel oil samples should be representative, taken according to ISO 3170 or ASTM D4057, and heated and homogenized before analysis because heavy fuel stratifies. Instrument temperature control and cleaning between samples determine repeatability: use solvent rinses and dry thoroughly. Flash point testers require barometric pressure input and periodic verification with reference fluids. XRF analyzers need warm-up time, calibration check samples and periodic film or window replacement. Centrifuges must be balanced and located on a stable bench. Because fuel oil is flammable and samples may contain hydrogen sulfide, use good ventilation, grounding and suitable personal protective equipment.
How do I select the right Micro Carbon Residue Tester for Fuel Oil?
Select a micro carbon residue tester when your laboratory must report carbon residue for ISO 8217 fuels and needs a smaller sample and shorter cycle than Conradson or Ramsbottom equipment. Specify number of positions, nitrogen supply, sample vial type and standards (ASTM D4530, ISO 10370).
How does it compare with other Fuel Oil Analysis variants?
Select instruments according to the specification you must certify. Marine fuel compliance to ISO 8217 needs density, sulfur, flash point, carbon residue, water, sediment and viscosity, so a laboratory normally combines several instruments. For sulfur, choose an ED-XRF analyzer if your limit is 0.1 to 3.5 percent, as with marine fuels and MARPOL limits, and UV fluorescence or WD-XRF only for trace levels. For flash point, choose manual or automatic Pensky-Martens testers, and note that the automatic version improves repeatability and safety. Density meters should have temperature control and viscosity correction for heavy fuel. Also evaluate throughput, sample heating, cleaning time, calibration standards, data-management features and whether the method is accepted by your customers or port authorities.
Full specifications
| Residue range | 0.01–30 % m/m |
|---|---|
| Final temperature | 500 °C |
| Temperature accuracy | ± 2 °C |
| Nitrogen flow | 600 ± 50 mL/min |
| Number of positions | 6 or 12 |
| Sample mass | ≈ 0.15–5 g (by residue level) |
| Cycle time | ≈ 30 min plus cooling |
| Repeatability | per ASTM D4530 |
| Power supply | 220 V AC, 50 Hz, 2 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Marine residual fuel qualification
- Crude oil and residue quality assessment
- Diesel fuel carbon residue on 10 % bottoms
Key features
- Heating block with precise 500 °C profile and ramp control
- Nitrogen flow control at 600 mL/min with mass flow controller
- Simultaneous testing of 6 to 12 sample vials
- Automatic cooling cycle under nitrogen atmosphere
- Built-in ASTM D4530 and ISO 10370 test programs
Frequently asked questions about Micro Carbon Residue Tester for Fuel Oil
Micro carbon residue and Conradson carbon residue measure similar behavior but use different apparatus. MCR results correlate closely and are reported as equivalent values, with a smaller sample and better repeatability.
Nitrogen prevents combustion so the sample cracks thermally rather than burning. A controlled nitrogen flow also sweeps away volatiles and keeps conditions consistent, which is essential for repeatable results.
Samples with high water or sediment content should be homogenized and possibly filtered, because uneven samples give poor repeatability. Foaming fuels require slower heating or smaller sample sizes according to the standard.
ISO 8217 lists density, kinematic viscosity, flash point, sulfur, carbon residue, water, sediment, ash, pour point and other parameters. Laboratories combine density meters, sulfur analyzers, flash point testers, viscometers and carbon residue testers to assess compliance.
Sulfur in residual fuel is usually measured by ED-XRF following ISO 8754 or ASTM D4294. The fuel is placed in a cell with a thin film, irradiated with X-rays and sulfur fluorescence intensity is compared with calibration standards.
Flash point indicates the temperature at which fuel vapors can ignite, so it defines safe storage and handling conditions. SOLAS requires marine fuels to have a minimum flash point of 60 degrees Celsius for safety.
Related models