Dual-Furnace Carbon Sulfur Analyzer
双炉碳硫分析仪
- SKU CSA-0003
- Carbon Sulfur Analyzers
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Carbon range | 0.0001–30% |
|---|---|
| Sulfur range | 0.0001–10% |
| Induction furnace sample mass | 0.2–1 g |
| Resistance furnace sample mass | 0.05–0.5 g |
| Induction analysis time | 30–60 s |
| Resistance analysis time | 2–5 min |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM E1019, ASTM E1941, ISO 15350
What is the Dual-Furnace Carbon Sulfur Analyzer?
A dual-furnace carbon sulfur analyzer combines an induction furnace and a resistance furnace in one system. It lets one laboratory analyze metals and non-metals without purchasing two instruments and sharing a common infrared detector set. It suits contract and multi-product laboratories.
What is the Dual-Furnace Carbon Sulfur Analyzer?
A dual-furnace carbon sulfur analyzer combines an induction furnace and a resistance furnace in one system. It lets one laboratory analyze metals and non-metals without purchasing two instruments and sharing a common infrared detector set. It suits contract and multi-product laboratories.
Overview
The dual-furnace carbon sulfur analyzer integrates both heating technologies, sharing the gas purification train and infrared cells through automated valve switching. Metal samples are analyzed in the induction furnace in about 40 seconds, while ores, cement, and other mineral powders are combusted in the tube furnace. Software selects the correct furnace and calibration set by method, which avoids mixing results from different matrices. Detector configuration usually includes separate high and low carbon cells and a sulfur cell, and the system can be fitted with a catalytic furnace for complete conversion of CO and SO3. The benefit is versatility for contract laboratories and multi-material manufacturers, while the trade-off is greater size and more complex maintenance because two heating systems and additional valves must be kept in service.
How does it work?
A carbon sulfur analyzer places a weighed sample of 0.05–1 g in a ceramic crucible, adds tungsten or iron and tin accelerators, and heats it in a high-frequency induction furnace or a resistance tube furnace in flowing oxygen. The accelerators couple with the induction field, melting the sample at 1500–2000 °C and oxidizing carbon to CO2 and sulfur to SO2. Dust is removed by a filter, and the gas passes through infrared absorption cells tuned to about 4.26 µm for CO2 and 7.35 µm for SO2. Absorbance follows the Beer–Lambert relationship, and integrated peak area is converted to mass percent using calibration with certified steel or ore standards. High-carbon and low-carbon channels with different path lengths extend dynamic range from ppm levels to several percent.
What are the key features?
- Induction and resistance furnaces share one set of infrared detectors.
- Method selection switches furnace, calibration, and parameters automatically.
- Metals analyzed in about 40 seconds and minerals in 2–5 minutes.
- Separate calibration sets prevent cross-matrix result contamination.
- Catalytic conversion ensures complete CO and SO3 oxidation in both furnaces.
- Consolidated data management records furnace type per sample for traceability.
What are the technical specifications?
A typical dual-furnace carbon sulfur analyzer configuration offers carbon range of 0.0001–30%, sulfur range of 0.0001–10%, induction furnace sample mass of 0.2–1 g, resistance furnace sample mass of 0.05–0.5 g, induction analysis time of 30–60 s and resistance analysis time of 2–5 min. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Dual-Furnace Carbon Sulfur Analyzer is used in contract testing laboratories with mixed matrices, steel plants with raw material labs, research laboratories in materials science and multi-product foundry and cement groups. Across the wider Carbon Sulfur Analyzers class, laboratories also deploy it for steel and cast iron composition control, ferroalloy and metal powder testing and ore, cement, and mineral carbon-sulfur content.
Which standards and certifications apply?
Relevant standards and methods include ASTM E1019, ASTM E1941, ISO 15350, ISO 9556, GB/T 20123, GB/T 223.69 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Crucibles must be pre-ignited at about 1000–1100 °C for one to two hours to reduce carbon blanks, and stored in a desiccator. Oxygen of at least 99.5% purity and a regulated supply of 2–3 bar are standard, with compressed air or nitrogen for pneumatic functions. The dust filter, anhydrone or magnesium perchlorate scrubbers, and furnace chamber need cleaning after a defined number of analyses, typically 30–100 for high-dust steels. Calibrate with certified reference materials bracketing the working range and run drift checks every few hours. Extraction of exhaust gases is required because SO2 is released during combustion. Induction furnaces emit radio-frequency energy and should be installed according to manufacturer guidance and IEC 61010-1 safety requirements.
How do I select the right Dual-Furnace Carbon Sulfur Analyzer?
Select the dual-furnace system when a single laboratory regularly analyzes both metals and minerals and bench space is limited. If 90% of samples are one type, a single-furnace model is cheaper. In an RFQ, define matrices, ranges, throughput per matrix, and whether simultaneous operation of both furnaces is required.
How does it compare with other Carbon Sulfur Analyzers variants?
Choose the furnace type first. High-frequency induction furnaces reach the temperatures needed for refractory metals, alloys, and ferroalloys, and give fast 30–60 second analyses. Resistance tube furnaces suit ores, cement, coal, and minerals, which do not couple with induction, and provide better stability for sulfur in non-metals. Next define the range: ppm-level carbon in ultra-low-carbon steel or titanium requires low-range cells, low blank, and careful crucible handling, while cast iron needs a high-range channel. Consider throughput through autosamplers and automatic cleaning, and check accessory needs such as crucible ovens, balances, and gas purifiers. Confirm that vendor documentation references ASTM E1019 or GB/T 20123, and ask for repeatability on certified reference materials representing your actual grades.
Full specifications
| Carbon range | 0.0001–30% |
|---|---|
| Sulfur range | 0.0001–10% |
| Induction furnace sample mass | 0.2–1 g |
| Resistance furnace sample mass | 0.05–0.5 g |
| Induction analysis time | 30–60 s |
| Resistance analysis time | 2–5 min |
| Furnace temperature | Induction 1500–2000 °C; tube up to 1350 °C |
| Detector | Infrared cells, high and low range |
| Oxygen supply | ≥99.5%, 2–3 bar |
| Power requirement | 220 V, 50 Hz, about 4 kW |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Contract testing laboratories with mixed matrices
- Steel plants with raw material labs
- Research laboratories in materials science
- Multi-product foundry and cement groups
Key features
- Induction and resistance furnaces share one set of infrared detectors.
- Method selection switches furnace, calibration, and parameters automatically.
- Metals analyzed in about 40 seconds and minerals in 2–5 minutes.
- Separate calibration sets prevent cross-matrix result contamination.
- Catalytic conversion ensures complete CO and SO3 oxidation in both furnaces.
- Consolidated data management records furnace type per sample for traceability.
Frequently asked questions about Dual-Furnace Carbon Sulfur Analyzer
Most dual-furnace systems share a detector train, so analyses run sequentially rather than simultaneously. Confirm this with the supplier if your workflow requires parallel operation, in which case two separate analyzers may be more appropriate.
Not inherently. Each furnace is calibrated with matrix-appropriate standards, and the shared infrared cells are the same as in single-furnace models. Accuracy depends on calibration, blank control, and maintenance rather than on the dual configuration.
The software stores a separate method and calibration for each matrix and furnace, and records the furnace used for each sample. This prevents applying a steel calibration to an ore sample and preserves traceability for audits.
It measures total carbon and total sulfur in solids such as steel, cast iron, alloys, ores, cement, and ceramics. The sample is burned in oxygen, and the resulting CO2 and SO2 are detected by infrared absorption cells, giving results from ppm levels to several percent.
Induction furnaces heat metals by electromagnetic coupling to 1500–2000 °C in seconds and suit steels and alloys. Resistance furnaces use heated tubes at 1200–1400 °C and handle non-conductive ores, cement, and minerals. The choice follows sample type.
Accelerators such as tungsten, iron chips, or tin increase induction coupling, raise combustion temperature, and promote complete, rapid oxidation of carbon and sulfur. Correct accelerator mass and purity matter because their carbon blank limits the lowest measurable concentration.