Manual Static-Jacket Bomb Calorimeter

手动静态水夹套氧弹量热仪

  • SKU CAL-0001
  • Calorimeters
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)

Price on request — FOB / CIF quoted per order

Prices depend on configuration, quantity and destination port.

Key specifications

Key specifications
Calorific value range5–35 MJ/kg typical for coal
Water equivalentAbout 10,000–15,000 J/K
Oxygen charging pressure25–30 bar
Sample mass0.5–1.2 g
Temperature resolution0.001 °C
Repeatability≤ 0.2 % RSD manual operation
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D5865, ISO 1928, GB/T 213

What is the Manual Static-Jacket Bomb Calorimeter?

A manual static-jacket bomb calorimeter burns a coal pellet in oxygen inside a fixed-temperature water jacket, and the operator reads the temperature rise to obtain the calorific value. It suits occasional testing, teaching and laboratories with limited budgets.

What is the Manual Static-Jacket Bomb Calorimeter?

A manual static-jacket bomb calorimeter burns a coal pellet in oxygen inside a fixed-temperature water jacket, and the operator reads the temperature rise to obtain the calorific value. It suits occasional testing, teaching and laboratories with limited budgets.

Overview

In this design the water jacket around the bomb is filled manually and left at ambient temperature, so the jacket drifts slightly during the test and the operator applies a radiation correction calculated from the temperature-time curve. Samples are weighed to about 1 g, pressed into a pellet and placed in the bomb with a known length of fuse wire and 25 to 30 bar of oxygen. The operator records temperatures at fixed intervals, reads the initial, main and after periods, and calculates the result or enters the readings into software performing the same calculation. Water equivalent is determined by burning certified benzoic acid. Repeatability is typically within about 0.2 % for a competent operator, sufficient for routine grading and instruction, but too slow and operator-dependent for a busy commercial laboratory.

How does it work?

A bomb calorimeter burns a pressed pellet of coal, about 0.5 to 1 g, in a high-pressure oxygen bomb immersed in a measured volume of water. The heat released raises the water temperature, and the temperature rise multiplied by the effective heat capacity of the system, called the water equivalent, gives the energy released. Corrections are applied for fuse wire, acid formation and the mass of the crucible contents. In an isoperibol design the jacket stays at a fixed temperature and the temperature-time curve is computed, in an adiabatic design the jacket follows the bomb temperature so heat loss is nearly zero, and in a manual static design the operator reads temperatures and applies a radiation correction. A microprocessor logs the curve, derives the water equivalent from certified benzoic acid and reports the result in MJ/kg.

What are the key features?

  • Fixed-temperature water jacket with manual temperature reading
  • Pressed pellet sample holder with fuse wire ignition
  • Oxygen bomb rated to about 30 bar working pressure
  • Water equivalent determined with certified benzoic acid
  • Software entry for temperature readings and radiation correction
  • Digital thermometer with 0.001 °C resolution

What are the technical specifications?

A typical manual static-jacket bomb calorimeter configuration offers calorific value range of 5–35 MJ/kg typical for coal, water equivalent of About 10,000–15,000 J/K, oxygen charging pressure of 25–30 bar, sample mass of 0.5–1.2 g, temperature resolution of 0.001 °C and repeatability of ≤ 0.2 % RSD manual operation. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Manual Static-Jacket Bomb Calorimeter is used in routine coal grading in small laboratories, teaching and operator training and blending checks and deposit sampling. Across the wider Calorimeters class, laboratories also deploy it for coal calorific value for pricing, power plant boiler efficiency monitoring and cement and steel kiln fuel checks.

Which standards and certifications apply?

Relevant standards and methods include ASTM D5865, ISO 1928, GB/T 213, ASTM D3175, ISO 18125, ASTM E711 and IEC 61010-1. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Install the calorimeter on a stable bench away from direct sunlight, draughts and heat sources, because ambient stability affects jacket temperature. Use oxygen of the specified purity from a regulator fitted with a pressure relief, and inspect the bomb, seals and valves on a schedule. Verify the water equivalent at the start of each working session with certified benzoic acid and monitor it as a trend, since drift indicates a leaking bomb, contaminated water or a failing sensor. Keep records of calibration and reference coal results, and compare them with the tolerances stated in the method. Confirm that report basis, sulfur and moisture corrections in the software match your contract.

How do I select the right Manual Static-Jacket Bomb Calorimeter?

Select the manual static-jacket unit when sample volume is low, budget is limited and an experienced operator is available, because accuracy depends on consistent technique. Specify bomb volume and pressure rating, thermometer resolution, water-equivalent range, and whether calculation software and certified benzoic acid are included.

How does it compare with other Calorimeters variants?

Selection turns on required accuracy, sample throughput and how much operator attention your laboratory can accept. Manual static-jacket units are the least expensive and are adequate for occasional checks, but they depend on the operator for temperature reading and correction. Isoperibol instruments dominate routine coal laboratories because they combine good accuracy with automatic temperature logging. Adiabatic calorimeters give the highest precision for reference work and for samples that burn slowly, such as some cokes and biomass, but they cost more and need careful setup. Check water-equivalent stability, reported resolution, the type of oxygen filling system, availability of benzoic acid and certified reference coal, and whether the software reports on the as-received, dry or dry ash-free basis your contract expects.

Full specifications

Full specifications
Calorific value range5–35 MJ/kg typical for coal
Water equivalentAbout 10,000–15,000 J/K
Oxygen charging pressure25–30 bar
Sample mass0.5–1.2 g
Temperature resolution0.001 °C
Repeatability≤ 0.2 % RSD manual operation
Test duration15–20 min
Jacket volumeAbout 2000 mL
Power supply220 V, 50/60 Hz

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Routine coal grading in small laboratories
  • Teaching and operator training
  • Blending checks and deposit sampling

Key features

  • Fixed-temperature water jacket with manual temperature reading
  • Pressed pellet sample holder with fuse wire ignition
  • Oxygen bomb rated to about 30 bar working pressure
  • Water equivalent determined with certified benzoic acid
  • Software entry for temperature readings and radiation correction
  • Digital thermometer with 0.001 °C resolution

Frequently asked questions about Manual Static-Jacket Bomb Calorimeter

The temperature rise of the water is multiplied by the water equivalent of the system to give the heat released, then corrected for fuse wire, acid formation and sample mass. Results are usually reported as gross calorific value at constant volume.