Pilot Airlift Bioreactor
中试气升式生物反应器
- SKU PIB-0003
- Pilot Bioreactors
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
- Pressure vessel compliance (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Working volume | 50–500 L |
|---|---|
| Height-to-diameter ratio | 5:1 to 10:1 |
| Material | 316L stainless steel |
| Aeration rate | 0.5–2 vvm |
| Temperature control | Jacket or internal coil |
| Design pressure | 0.1–0.3 MPa |
By Lab Manufacturers China Technical TeamLast reviewed Standards: EU GMP Annex 1, EU GMP Annex 11, 21 CFR Part 11
What is the Pilot Airlift Bioreactor?
A pilot airlift bioreactor circulates the culture with gas in a tall column, providing low-shear mixing and good oxygen transfer. It suits shear-sensitive cells, viscous broths and filamentous organisms at pilot scale.
What is the Pilot Airlift Bioreactor?
A pilot airlift bioreactor circulates the culture with gas in a tall column, providing low-shear mixing and good oxygen transfer. It suits shear-sensitive cells, viscous broths and filamentous organisms at pilot scale.
Overview
A pilot airlift bioreactor applies gas-driven circulation at intermediate scale, where its low shear and simple internals become significant advantages. Gas injected at the base of the riser lowers local density, so liquid rises there and descends through the downcomer, creating continuous circulation without an impeller. An internal draft tube or an external loop defines the two zones, and the tall column gives a long gas-liquid contact path and efficient oxygen transfer. Because there are no shaft seals or impellers in the broth, maintenance, cleaning and sterilization are simpler. Temperature is controlled by a jacket or coil, and instrumentation covers pH, dissolved oxygen, foam and pressure. The design suits filamentous fungi, plant and animal cells and viscous broths, though building height and foam control are the main constraints.
How does it work?
A pilot bioreactor applies the same biological principles as a smaller vessel but at a scale where mixing, oxygen transfer and heat removal behave differently. Larger impellers and higher gas flow increase oxygen transfer, described by the volumetric mass-transfer coefficient, while a jacket or coil removes the greater metabolic heat load. Sensors for temperature, pH, dissolved oxygen, pressure and often foam or level sit in hygienic fittings, and cascade control adjusts agitation, gas composition and back pressure to hold dissolved oxygen. Because mixing time, tip speed and mass-transfer coefficient cannot all be matched to bench scale, engineers choose the most critical criterion and verify performance during the run. Pilot vessels may be reusable stainless steel or single-use, and they include defined sampling, feed and harvest points.
What are the key features?
- Gas-driven circulation with riser and downcomer
- Internal draft tube or external loop
- Low shear with no seals in the broth
- Tall column for efficient oxygen transfer
- Sterilize-in-place and clean-in-place design
What are the technical specifications?
A typical pilot airlift bioreactor configuration offers working volume of 50–500 L, height-to-diameter ratio of 5:1 to 10:1, material of 316L stainless steel, aeration rate of 0.5–2 vvm, temperature control of Jacket or internal coil and design pressure of 0.1–0.3 MPa. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Pilot Airlift Bioreactor is used in filamentous fungal fermentation, shear-sensitive cell culture, viscous broth production and scale-up of low-shear processes. Across the wider Pilot Bioreactors class, laboratories also deploy it for scale-up verification, clinical and trial batch production and monoclonal antibody and protein production.
Which standards and certifications apply?
Relevant standards and methods include EU GMP Annex 1, EU GMP Annex 11, 21 CFR Part 11, 21 CFR Part 211, GB/T 150, ISO 13485, USP <87>, USP <88>, ICH Q11 and IEC 61010-1. Certification candidates for this model are CE, ISO 9001 manufacturing and Pressure vessel compliance; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Qualify the vessel and its utilities with documented pressure, temperature, cleaning and sterilization tests. Calibrate pH and dissolved oxygen probes before each run and verify sensor placement so readings reflect bulk conditions. Establish the scale-up criterion and document why it was chosen, then record the supporting data. Monitor foam, heat removal and dissolved oxygen closely, since these are the usual scale-dependent problems. Keep batch records that capture setpoints, deviations and offline measurements such as cell count and metabolite levels. For reusable vessels, control cleaning cycles and gaskets; for single-use, manage bag lots, integrity and extractables data. Confirm electrical safety to IEC 61010-1.
How do I select the right Pilot Airlift Bioreactor?
Choose a pilot airlift for shear-sensitive or viscous cultures and where low maintenance is valued. In an RFQ specify volume, height and required headroom, loop design, aeration and enrichment capability, temperature control and foam management features.
How does it compare with other Pilot Bioreactors variants?
Size the vessel to reproduce the most important bench parameter and to produce enough material for the next step, whether that is a clinical batch, a trial quantity or data for production design. Decide between stainless steel, which suits repeated campaigns and higher pressure, and single-use, which shortens changeover and reduces cleaning validation. Match the geometry, impeller type and gas handling to the scale-up criterion you intend to hold constant. List control loops and ranges, the number of feed, sampling and harvest ports, and whether redox, off-gas analysis or back pressure control are needed. Evaluate clean-in-place and sterilize-in-place for reusable designs, single-use bag performance for disposable designs, and the records and qualification package available for regulated work.
Full specifications
| Working volume | 50–500 L |
|---|---|
| Height-to-diameter ratio | 5:1 to 10:1 |
| Material | 316L stainless steel |
| Aeration rate | 0.5–2 vvm |
| Temperature control | Jacket or internal coil |
| Design pressure | 0.1–0.3 MPa |
| Dissolved oxygen control | Via gas flow, enrichment or back pressure |
| Foam control | Mechanical breaker and antifoam |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Filamentous fungal fermentation
- Shear-sensitive cell culture
- Viscous broth production
- Scale-up of low-shear processes
Key features
- Gas-driven circulation with riser and downcomer
- Internal draft tube or external loop
- Low shear with no seals in the broth
- Tall column for efficient oxygen transfer
- Sterilize-in-place and clean-in-place design
Frequently asked questions about Pilot Airlift Bioreactor
Filamentous fungi, plant and animal cells and other shear-sensitive organisms benefit because there are no impellers to damage them. Very high oxygen demand or highly viscous broths may still favor a stirred tank with oxygen enrichment.
The tall column requires substantial building height, and foam control is critical because gas drives both aeration and circulation. Oxygen transfer can become limiting at very high cell densities, when enrichment or higher back pressure is needed.
Circulation follows the gas flow rate, which changes the density difference between riser and downcomer and therefore liquid velocity. Dissolved oxygen control adjusts gas flow, oxygen fraction or back pressure rather than an impeller speed.
It verifies that a process developed at bench scale behaves as expected at larger volume and it produces enough material for clinical trials, product evaluation or production design. It also reveals scale-dependent effects such as oxygen limitation or poor heat removal.
Constant volumetric mass-transfer coefficient, tip speed, power per volume and mixing time cannot all be held simultaneously. The team selects the criterion most related to product quality or yield, applies it at pilot scale and verifies the result before transferring to production.
Stainless steel suits repeated campaigns, higher pressures and steam sterilization, while single-use suits multi-product or clinical work where cleaning validation between products would be costly. Extractables data, volume and pressure limits and supply continuity guide the single-use choice.