Insulating Oil Oxidation Stability Tester
绝缘油氧化安定性测定仪
- SKU OXS-0005
- Oxidation Stability
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Test temperature | 110 or 120 °C ± 0.5 °C |
|---|---|
| Air or oxygen flow | 15 ± 1 L/h per tube |
| Sample mass | 25 ± 0.1 g |
| Test duration | 164–500 h |
| Tube positions | 8–16 |
| Catalyst | Copper wire coil |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D2272, ASTM D943, ASTM D6186
What is the Insulating Oil Oxidation Stability Tester?
An insulating oil oxidation stability tester ages transformer oil with air or oxygen and copper catalyst at 110 to 120 °C, then measures acid number and sludge formed. It is used to classify inhibited and uninhibited mineral insulating oils under IEC 61125 and ASTM D2440.
What is the Insulating Oil Oxidation Stability Tester?
An insulating oil oxidation stability tester ages transformer oil with air or oxygen and copper catalyst at 110 to 120 °C, then measures acid number and sludge formed. It is used to classify inhibited and uninhibited mineral insulating oils under IEC 61125 and ASTM D2440.
Overview
A 25 g oil sample in a glass tube with a copper catalyst coil is heated in an aluminum block or bath at 120 °C while dry air or oxygen is bubbled through at 15 L per hour for 164 to 500 hours, depending on oil class. After aging, the oil is analyzed for volatile acid, total acid and sludge. IEC 61125 method C uses 120 °C and 164 hours for uninhibited oils. Condensers collect volatile acids for titration. Multi-position blocks run eight to sixteen tubes together and control flow for each tube with needle valves and flow meters. The test shows how well an oil resists forming acid and sludge that degrade paper insulation and cooling efficiency in transformers.
How does it work?
Oxidation converts hydrocarbons into acids, sludge and varnish through free-radical chain reactions that speed up once inhibitors are consumed. Accelerated tests expose an oil sample to oxygen or air at elevated temperature, often with copper and iron catalyst wire and water, and monitor the reaction. In rotating pressure vessel tests, a sealed bomb containing oil, water and a copper coil is charged with oxygen at 620 kPa, heated to 150 °C and rotated, and the time until pressure falls by 175 kPa is recorded. Turbine oil tests run for thousands of hours and measure time to an acid number of 2.0 mg KOH/g. Differential scanning calorimetry detects the exothermic onset under oxygen pressure, and thin-film methods measure oxygen uptake. Induction time reflects the oil's remaining antioxidant reserve.
What are the key features?
- Heating block for 8 to 16 aging tubes
- Individual flow control and flow meters
- Condensers with volatile acid trap
- Temperature control at 110–120 °C
- Timer and alarm for 164–500 h tests
What are the technical specifications?
A typical insulating oil oxidation stability tester configuration offers test temperature of 110 or 120 °C ± 0.5 °C, air or oxygen flow of 15 ± 1 L/h per tube, sample mass of 25 ± 0.1 g, test duration of 164–500 h, tube positions of 8–16 and catalyst of Copper wire coil. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Insulating Oil Oxidation Stability Tester is used in transformer oil qualification, incoming insulating oil inspection and oxidation inhibitor evaluation. Across the wider Oxidation Stability class, laboratories also deploy it for turbine oil life assessment, antioxidant additive development and hydraulic oil QC.
Which standards and certifications apply?
Relevant standards and methods include ASTM D2272, ASTM D943, ASTM D6186, ASTM D4742, ASTM D7545, IEC 61125, ASTM D2440 and ISO 4263. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Oxidation tests use pressurized oxygen and heated oil, so safety is the first consideration. Use oxygen-compatible, oil-free regulators and fittings, pressure relief devices and guarding, and keep the area ventilated and free of grease. Clean glassware and bombs thoroughly, since contaminants alter induction time. Prepare catalyst coils exactly per the method, using specified copper wire and polishing. Check bath temperature uniformity and sensor calibration, and test with a reference oil of known RPVOT value on a schedule. Verify leak tightness before each run and keep logs of repeat results. Dispose of used oil and water according to local rules.
How do I select the right Insulating Oil Oxidation Stability Tester?
Select this tester for insulating oil production and acceptance testing where IEC 61125 or ASTM D2440 is specified. Specify number of tubes, temperature, gas type, condenser design and whether titration equipment for acid and sludge is required. It is not needed for turbine or engine oils.
How does it compare with other Oxidation Stability variants?
Select the test that your specification names. Turbine and hydraulic oil specifications commonly call for rotating pressure vessel oxidation test (RPVOT) and the long-duration turbine oil stability test. Formulators developing additives prefer faster methods such as pressure differential scanning calorimetry or thin-film oxygen uptake, which need small samples. Insulating oil producers follow IEC 61125 or ASTM D2440. Consider the number of positions, since tests run for hours or weeks, and the degree of automation including pressure logging, endpoint detection and report generation. Check bath temperature uniformity, safety features for oxygen and pressure, and supply of glass sample containers, catalyst coil and test bombs. In the RFQ, list methods, number of simultaneous tests and sample volumes.
Full specifications
| Test temperature | 110 or 120 °C ± 0.5 °C |
|---|---|
| Air or oxygen flow | 15 ± 1 L/h per tube |
| Sample mass | 25 ± 0.1 g |
| Test duration | 164–500 h |
| Tube positions | 8–16 |
| Catalyst | Copper wire coil |
| Temperature uniformity | ± 0.5 °C |
| Gas supply | Dry air or oxygen |
| Applicable methods | IEC 61125, ASTM D2440 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Transformer oil qualification
- Incoming insulating oil inspection
- Oxidation inhibitor evaluation
Key features
- Heating block for 8 to 16 aging tubes
- Individual flow control and flow meters
- Condensers with volatile acid trap
- Temperature control at 110–120 °C
- Timer and alarm for 164–500 h tests
Frequently asked questions about Insulating Oil Oxidation Stability Tester
Inhibited oils contain antioxidant additives such as DBPC, so they resist oxidation for longer. IEC 60296 classifies oils as inhibited, uninhibited or trace-inhibited by oxidation stability limits.
Sludge deposits reduce heat dissipation and can clog cooling ducts in transformers. Acid attacks cellulose insulation, so low acid and sludge values indicate longer transformer life.
Duration depends on oil class, for example 164 hours for uninhibited oils at 120 °C and up to 500 hours for inhibited oils at 120 °C. Follow the method and oil specification.
It is the resistance of the oil to chemical reaction with oxygen, which forms acids, sludge and varnish. Testing measures how long the oil lasts under accelerated conditions before it breaks down.
RPVOT measures the minutes until oxygen pressure in a heated, rotating bomb with water and copper catalyst falls by 175 kPa. A longer time indicates greater remaining antioxidant reserve in the oil.
RPVOT takes from tens of minutes to a few hundred minutes. Pressure DSC takes minutes to an hour, while turbine oil stability tests can last thousands of hours.
Related models