Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust
带顶罩排风的 II 级 A2 型细胞毒性药物安全柜
- SKU CYC-0003
- Cytotoxic Cabinets
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Cabinet type | Class II Type A2 with canopy |
|---|---|
| Inflow velocity | 0.38–0.51 m/s |
| Downflow velocity | 0.25–0.35 m/s |
| Recirculation | ≈ 70 % |
| Exhaust | 30 % via canopy duct |
| Canopy type | Non-sealing, thimble |
By Lab Manufacturers China Technical TeamLast reviewed Standards: EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), USP <800> (hazardous drugs handling in healthcare settings)
What is the Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust?
A Class II Type A2 cytotoxic cabinet with canopy exhaust recirculates most air but captures the exhaust in an outdoor duct. It suits lower-risk drugs where full recirculation control is not required.
What is the Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust?
A Class II Type A2 cytotoxic cabinet with canopy exhaust recirculates most air but captures the exhaust in an outdoor duct. It suits lower-risk drugs where full recirculation control is not required.
Overview
This cabinet uses the A2 airflow pattern, recirculating about 70 % of HEPA-filtered air and exhausting 30 %, but a canopy hood over the exhaust collects that fraction and carries it outdoors through a duct. The canopy is a partial enclosure that does not seal to the cabinet, so it does not affect cabinet balance and certification. Inflow remains 0.38–0.51 m/s and downflow 0.25–0.35 m/s. The arrangement reduces room contamination from the 30 % exhaust and odours, but because most air recirculates, it is not suitable for volatile hazardous drugs or large aerosol loads. Widths of 1200–1800 mm are common. It is used where a B2 installation is not feasible and the pharmacy risk assessment permits an A2 cabinet with canopy for the drugs in question.
How does it work?
Cytotoxic cabinets combine the airflow of a Class II biosafety cabinet with a design that keeps hazardous drug particles and vapours away from the operator and out of the room. Room air enters at the front grille, forming an air curtain across the opening, and HEPA-filtered air flows downward over the work zone to protect the preparation. Exhaust is HEPA filtered, and the cabinet is typically operated without recirculation into the work zone, with all exhaust sent to a dedicated duct. Many designs hold the work surface at negative pressure and use a split or closed front so drug aerosols cannot escape. Some pharmacy units use sealed glove ports, giving isolator-style containment with no direct operator contact. Surfaces are smooth stainless steel or plastic for frequent decontamination, and the cabinet is a classified environment for aseptic operations. Certification covers HEPA integrity, airflow and containment testing.
What are the key features?
- A2 airflow with 70 % recirculation and 30 % exhaust
- Canopy hood collecting exhaust to outdoor duct
- Inflow 0.38–0.51 m/s, downflow 0.25–0.35 m/s
- Canopy does not seal, so balance is unaffected
- Reduced room odour and contamination
- Widths 1200–1800 mm
What are the technical specifications?
A typical class ii type a2 cytotoxic cabinet with canopy exhaust configuration offers cabinet type of Class II Type A2 with canopy, inflow velocity of 0.38–0.51 m/s, downflow velocity of 0.25–0.35 m/s, recirculation of ≈ 70 %, exhaust of 30 % via canopy duct and canopy type of Non-sealing, thimble. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust is used in lower-risk hazardous drug prep, pharmacies without B2 exhaust, aseptic compounding with canopy and clinical and satellite pharmacies. Across the wider Cytotoxic Cabinets class, laboratories also deploy it for antineoplastic drug compounding, hospital pharmacy aseptic preparation and hazardous drug dose preparation.
Which standards and certifications apply?
Relevant standards and methods include EN 12469 (biosafety cabinets), NSF/ANSI 49 (biosafety cabinetry), USP <800> (hazardous drugs handling in healthcare settings), USP <797> (pharmaceutical compounding, sterile preparations), ISO 14644-1 (cleanroom classification) and EU GMP Annex 1 (sterile manufacturing). Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Place the cabinet in a controlled area with correct room pressure relationships, commonly a negative-pressure buffer room, and away from doors and air diffusers. Change gloves and sleeves on a defined schedule and inspect them for pinholes, since operator protection depends on intact barrier and airflow. Decontaminate the work surface and interior after every session and use a closed-system transfer device where required. Verify HEPA integrity, inflow and containment at installation and at least annually, and after relocation or filter change. Do not exceed the maximum drug volume or use unapproved agents that would overwhelm the exhaust. Manage waste, sharps and spill kits according to the hazardous drug procedure, and keep a log of certification and decontamination.
How do I select the right Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust?
Choose the canopy A2 only where B2 is not feasible and the risk assessment permits it for the drugs used. Confirm canopy airflow and duct route do not disturb the cabinet. RFQ: drug list, canopy airflow, duct route, room pressure and certification.
How does it compare with other Cytotoxic Cabinets variants?
Select the cabinet from the hazardous drug list, the required dose accuracy and the pharmacy's aseptic requirements. Use a Class II Type B2 cytotoxic cabinet with 100 % dedicated exhaust when volatile hazardous drugs are prepared. Choose a Compounding Aseptic Containment Isolator where the highest operator protection is needed and full isolation is acceptable, since glove access reduces dexterity. A Type A2 cytotoxic cabinet with canopy exhaust suits lower-risk drugs when recirculation is limited and local rules allow it. Confirm compatibility with USP <800> and USP <797> requirements, the pharmacy's risk assessment and any local rules on where the cabinet is placed, including a negative-pressure buffer room. Verify certification, plan for HEPA leak and containment testing, and specify the transfer method for drugs and waste.
Full specifications
| Cabinet type | Class II Type A2 with canopy |
|---|---|
| Inflow velocity | 0.38–0.51 m/s |
| Downflow velocity | 0.25–0.35 m/s |
| Recirculation | ≈ 70 % |
| Exhaust | 30 % via canopy duct |
| Canopy type | Non-sealing, thimble |
| HEPA grade | H14 supply and exhaust |
| Exhaust airflow | 200–500 m3/h |
| Width | 1200–1800 mm |
| Work surface | Stainless steel 304 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Lower-risk hazardous drug prep
- Pharmacies without B2 exhaust
- Aseptic compounding with canopy
- Clinical and satellite pharmacies
Key features
- A2 airflow with 70 % recirculation and 30 % exhaust
- Canopy hood collecting exhaust to outdoor duct
- Inflow 0.38–0.51 m/s, downflow 0.25–0.35 m/s
- Canopy does not seal, so balance is unaffected
- Reduced room odour and contamination
- Widths 1200–1800 mm
Frequently asked questions about Class II Type A2 Cytotoxic Cabinet with Canopy Exhaust
No. It captures only the 30 % exhaust fraction, while bulk air still recirculates, so it does not remove all drug vapour. Full B2 exhaust is safer for hazardous drugs.
A small air gap or thimble connection lets the building exhaust draw air without pulling on the cabinet, so airflow balance and certification are unaffected. Sealing it could disturb containment.
Volatile hazardous drugs and any agent producing high vapour or aerosol loads, because they could build up in recirculated air. Use a B2 cabinet for those agents.
It is a Class II biosafety cabinet or isolator configured to contain hazardous drugs such as antineoplastics during compounding. It protects the operator from drug aerosols and the product from contamination, with dedicated exhaust and no work-zone recirculation.
A Class II Type B2 cabinet with 100 % dedicated exhaust, or a Compounding Aseptic Containment Isolator, is typically preferred. Both avoid recirculating drug vapour into the room and follow USP <800> guidance.
Because HEPA filters capture particles but not all drug vapour, and recirculated air could carry contamination back into the workspace. Dedicated exhaust to the outside removes both particles and vapour.