Mini Portable Thermal Cycler
迷你便携式热循环仪
- SKU COP-0005
- Conventional PCR
- CE (to be confirmed)
- ISO 9001 manufacturing (to be confirmed)
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
Key specifications
| Block format | 16, 32 or 48 × 0.2 mL wells |
|---|---|
| Temperature range | 4–99 °C |
| Max block ramp rate | 2–3 °C/s |
| Block uniformity | ±0.5 °C |
| Temperature accuracy | ±0.3 °C |
| Reaction volume | 10–50 µL |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ISO 20395, ISO 22174, IEC 61010-1
What is the Mini Portable Thermal Cycler?
A mini portable thermal cycler is a compact 16–48 well unit that weighs a few kilograms, often with a built-in touchscreen and optional battery or vehicle power. It is the right choice for teaching, field testing, small laboratories and low-volume work where bench space and sample count are limited.
What is the Mini Portable Thermal Cycler?
A mini portable thermal cycler is a compact 16–48 well unit that weighs a few kilograms, often with a built-in touchscreen and optional battery or vehicle power. It is the right choice for teaching, field testing, small laboratories and low-volume work where bench space and sample count are limited.
Overview
The mini cycler packs a small Peltier block, low-power controller and compact heated lid into a unit of about 2–4 kg. Common formats hold 16 or 32 tubes of 0.2 mL, or 48 wells for small plates. Ramp rates are around 2–3 °C/s and uniformity is about ±0.5 °C, slightly looser than full-size models but adequate for routine end-point PCR. Many units can be powered from a 12 V DC source or an external battery pack, allowing field work. Program storage is limited to tens of protocols, and USB export is standard. Because of the smaller block and fan, run noise is low, which suits teaching rooms. These units are not intended for high-throughput or strict quality-system use.
How does it work?
A thermal cycler holds reaction tubes or plates in a metal block whose temperature is controlled by Peltier thermoelectric elements. Each cycle denatures double-stranded DNA at about 94–98 °C, cools to the primer annealing temperature, typically 50–68 °C, and then heats to about 72 °C so a thermostable DNA polymerase extends the primers. Repeating this 25–40 times amplifies the target exponentially. Key components include the Peltier-driven block, a heat sink and fan, a temperature sensor feedback loop, a heated lid that prevents condensation and evaporation, and a controller that stores multi-step programs. Block uniformity and ramp rate determine how reproducibly each well reaches its set temperature.
What are the key features?
- Compact 2–4 kg body fits on crowded benches or in a field case
- 12 V DC or battery power option for field use
- Low-noise fan suited to teaching rooms
- Touchscreen with simple protocol wizard
- USB export of run logs and programs
- Heated lid with automatic height adjustment
What are the technical specifications?
A typical mini portable thermal cycler configuration offers block format of 16, 32 or 48 × 0.2 mL wells, temperature range of 4–99 °C, max block ramp rate of 2–3 °C/s, block uniformity of ±0.5 °C, temperature accuracy of ±0.3 °C and reaction volume of 10–50 µL. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
Which applications is it used for?
The Mini Portable Thermal Cycler is used in teaching and training labs, field-based pathogen screening, small research groups and low-volume end-point PCR. Across the wider Conventional PCR class, laboratories also deploy it for gene cloning and plasmid construction, genotyping and mutation screening and pathogen detection by end-point PCR.
Which standards and certifications apply?
Relevant standards and methods include ISO 20395, ISO 22174, IEC 61010-1, IEC 61010-2-010, MIQE guidelines (research reporting) and CLSI MM19. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install the cycler on a stable bench with at least 10 cm of clearance for air intake and exhaust, away from direct sunlight and drafts. Verify block temperature accuracy and uniformity at installation and at least annually using a calibrated multi-point probe, and record results for audit. Clean the block wells with a lint-free swab and 70% ethanol to remove spilled reagent and prevent cross-contamination, and inspect the heated lid seal. Use sealing films or caps matched to the lid pressure. Keep pre-PCR and post-PCR work areas physically separated, and run no-template controls to detect amplicon carry-over. Firmware updates and program backups should be controlled under the laboratory's document system.
How do I select the right Mini Portable Thermal Cycler?
Select a mini portable cycler for teaching, fieldwork or occasional PCR when bench space is tight. It costs less than full-size units but has looser uniformity and fewer programs. In an RFQ, specify tube count, weight, DC or battery power option, noise level, program memory, and whether calibration documentation is needed for regulated work. For diagnostics or strict quality systems, choose a full-size model.
How does it compare with other Conventional PCR variants?
Choose first by throughput and format: 96-well blocks suit most research work, 384-well blocks suit high-throughput screening, and small 16–48 well units suit teaching or occasional use. Next, compare ramp rate and block uniformity, because fast cyclers shorten run time but need polymerases and buffers validated for short steps. Dual-block instruments allow two independent programs at once and help shared core facilities. Check the heated lid range, programmable features such as touchdown and gradient, and whether the block is interchangeable. Consider connectivity for audit trails if the laboratory works under quality systems, and confirm the tube or plate formats you use, including low-profile plates and strip tubes.
Full specifications
| Block format | 16, 32 or 48 × 0.2 mL wells |
|---|---|
| Temperature range | 4–99 °C |
| Max block ramp rate | 2–3 °C/s |
| Block uniformity | ±0.5 °C |
| Temperature accuracy | ±0.3 °C |
| Reaction volume | 10–50 µL |
| Heated lid range | 30–110 °C |
| Program memory | ≥ 50 programs |
| Weight | 2–4 kg |
| Power supply | 100–240 V AC adapter, 12 V DC input option |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
Applications
- Teaching and training labs
- Field-based pathogen screening
- Small research groups
- Low-volume end-point PCR
Key features
- Compact 2–4 kg body fits on crowded benches or in a field case
- 12 V DC or battery power option for field use
- Low-noise fan suited to teaching rooms
- Touchscreen with simple protocol wizard
- USB export of run logs and programs
- Heated lid with automatic height adjustment
Frequently asked questions about Mini Portable Thermal Cycler
Some models accept a 12 V DC supply or external battery pack and can run for several hours, depending on protocol and capacity. Confirm the required battery rating and runtime with the supplier for your field protocol.
Uniformity is typically ±0.5 °C, adequate for many end-point assays but looser than full-size models. Validate each assay on the unit, and use a full-size calibrated cycler where regulations require tighter performance records.
Typical capacity is 16, 32 or 48 reactions per run. This suits teaching classes and small projects, but throughput is too low for routine diagnostics or large screens.
It amplifies a chosen DNA segment by cycling a reaction mix through denaturation, annealing and extension temperatures. The product is analysed after the run, usually on an agarose gel. Typical uses include cloning, genotyping, sequencing template preparation, and qualitative detection of pathogens or transgenes.
Conventional PCR reads the result at the end of the run, usually by gel electrophoresis, and gives a qualitative yes or no answer. Real-time PCR measures fluorescence every cycle and can quantify starting template. Conventional cyclers cost less and have no optical module.
Block temperature accuracy and uniformity should be verified at installation and at least once a year, or after any repair or relocation. Laboratories working under ISO 17025 or GMP often shorten this interval and keep calibration certificates traceable to national standards.