Key specifications

Key specifications
Chamber volume100–400 L
Temperature range0–60 C, set at 20 C
Temperature uniformity±0.5–1 C
Programs10–30
Logging interval1–60 min
Data interfaceRS-485, USB or Ethernet
Request a quote Download brochure · to be provided

By Lab Manufacturers China Technical TeamLast reviewed Standards: IEC 61010-1, IEC 61010-2-010, DIN 12880

What is the Programmable BOD Incubator?

A programmable BOD incubator records and controls the incubation profile, providing traceable temperature evidence for the five-day test. It is chosen where methods or accreditation require a documented temperature history for every test.

What is the Programmable BOD Incubator?

A programmable BOD incubator records and controls the incubation profile, providing traceable temperature evidence for the five-day test. It is chosen where methods or accreditation require a documented temperature history for every test.

Overview

This incubator keeps the fixed BOD setpoint but adds a multi-segment controller and continuous temperature logging. The controller stores programs with setpoints and durations, alarms on deviation, and records temperature with timestamps through RS-485, USB or Ethernet so a valid incubation can be demonstrated. Real-time and historical curves appear on the display, and the log can be exported for reports. Refrigerated cooling holds 20 C with uniformity of plus or minus 0.5 to 1 C, and defrost is managed to avoid excursions. Chamber volumes run 100 to 400 liters. It suits accredited environmental laboratories that must show temperature was maintained, and research work needing reproducible incubation histories.

How does it work?

A BOD incubator maintains a fixed setpoint, commonly 20 C, with compressor refrigeration and heating under PID control, so it holds temperature whether the room is warmer or cooler. Forced-air circulation keeps uniformity within plus or minus 0.5 to 1 C across the chamber, and a door heater or heated frame prevents condensation at the fixed low temperature. The interior is dark, since light can affect samples and encourage algae, and shelves or racks hold BOD bottles or other vessels. Defrost cycles are scheduled to avoid temperature excursions during the five-day test, and many models include a timer or chart for the incubation period. Alarms warn of temperature deviation, and some units log temperature continuously so a valid test can be demonstrated.

What are the key features?

  • Continuous temperature logging with timestamps
  • Multi-segment program storage
  • Deviation and over-temperature alarms
  • RS-485, USB or Ethernet data export
  • Real-time and historical curve display
  • Managed defrost to avoid excursions

What are the technical specifications?

A typical programmable bod incubator configuration offers chamber volume of 100–400 L, temperature range of 0–60 C, set at 20 C, temperature uniformity of ±0.5–1 C, programs of 10–30, logging interval of 1–60 min and data interface of RS-485, USB or Ethernet. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.

Which applications is it used for?

The Programmable BOD Incubator is used in accredited BOD testing with records, regulated wastewater monitoring, method validation and audits and research with reproducible incubation. Across the wider BOD Incubators class, laboratories also deploy it for five-day BOD determination in water and wastewater, biochemical oxygen demand method training and influent and effluent monitoring.

Which standards and certifications apply?

Relevant standards and methods include IEC 61010-1, IEC 61010-2-010, DIN 12880, APHA 5210, ISO 5815-1 and ISO 9001. Certification candidates for this model are CE, ISO 9001 manufacturing and RoHS; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.

Place the incubator in a ventilated location with clearance for the condenser and door swing, away from heat sources and direct sunlight. Keep the interior dark and do not open the door during the incubation other than at defined checkpoints, since temperature recovery may alter the result. Verify the temperature at the working setpoint with a calibrated probe, and log it continuously where the method requires evidence. Clean spills promptly and disinfect as needed, taking care not to scratch the interior or leave residues that affect samples. Check the condenser, gaskets and door heater during scheduled service, and record all maintenance.

How do I select the right Programmable BOD Incubator?

Select the programmable BOD incubator when the laboratory must demonstrate temperature throughout the incubation, as for accreditation or client audits. Standard units are cheaper where records are not needed. In the RFQ, specify logging interval, data interface, record format, alarms and chamber volume.

How does it compare with other BOD Incubators variants?

Choose a BOD incubator when a method specifies incubation at a fixed low temperature, most often the 20 C five-day BOD test, and when the laboratory needs to show that temperature was held. Set the required temperature and tolerance, then check uniformity at that point rather than a midpoint. Match chamber volume to bottle count and to the use of racks or shelves. Confirm defrost behaviour, since ice build-up or regular defrost cycles can disturb the incubation. Decide whether programmable control, dual chambers or temperature logging are needed for compliance. Check door type, alarms, data output and refrigerant type before ordering.

Full specifications

Full specifications
Chamber volume100–400 L
Temperature range0–60 C, set at 20 C
Temperature uniformity±0.5–1 C
Programs10–30
Logging interval1–60 min
Data interfaceRS-485, USB or Ethernet
AlarmsDeviation and over-temperature
Cooling systemCompressor
DefrostScheduled automatic
Power supply220 V, 50/60 Hz

Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.

Applications

  • Accredited BOD testing with records
  • Regulated wastewater monitoring
  • Method validation and audits
  • Research with reproducible incubation

Key features

  • Continuous temperature logging with timestamps
  • Multi-segment program storage
  • Deviation and over-temperature alarms
  • RS-485, USB or Ethernet data export
  • Real-time and historical curve display
  • Managed defrost to avoid excursions

Frequently asked questions about Programmable BOD Incubator

A valid five-day BOD result depends on temperature being held at 20 C, so accredited and client-driven work requires evidence. Continuous logging with timestamps shows whether any excursion occurred and supports the result during audits or disputes.