
Manual Kinematic Viscosity Bath with Glass Capillary Viscometers
手动运动黏度恒温浴及玻璃毛细管黏度计
- CE (to be confirmed)
- +1
Welcome to Lab Manufacturers China — laboratory instruments direct from a Chinese manufacturer

冷启动模拟器
Price on request — FOB / CIF quoted per order
Prices depend on configuration, quantity and destination port.
| Temperature range | −35 to 0 °C |
|---|---|
| Temperature stability | ± 0.05 °C |
| Viscosity range | 500–10,000 mPa·s |
| Shear rate | ≈ 10⁵–10⁶ s⁻¹ |
| Sample volume | ≈ 0.25–0.5 mL |
| Rotor speed range | ≈ 50–300 rpm |
By Lab Manufacturers China Technical TeamLast reviewed Standards: ASTM D445, ISO 3104, ASTM D2270
A cold cranking simulator measures apparent viscosity of engine oils at low temperatures, about −10 to −35 °C, at the high shear rate of a cold engine start. It supports SAE J300 W-grade classification of passenger-car and heavy-duty engine oils to ASTM D5293.
A cold cranking simulator measures apparent viscosity of engine oils at low temperatures, about −10 to −35 °C, at the high shear rate of a cold engine start. It supports SAE J300 W-grade classification of passenger-car and heavy-duty engine oils to ASTM D5293.
The instrument contains a cooled stator with a close-fitting rotor driven by a motor. A small oil sample fills the gap, and the rotor turns at a speed that depends on viscosity under constant applied voltage. Calibration oils of known viscosity at each test temperature create a calibration curve relating rotor speed to apparent viscosity in mPa·s. Shear rate is about 10^5 to 10^6 reciprocal seconds, similar to that in starter-motor cranking. Temperature control uses a refrigerated circulator or thermoelectric system, and sample temperature must be held within ± 0.1 °C. Results are compared with the limits for grades such as 0W, 5W and 10W in SAE J300. A single measurement takes a few minutes once the stator has stabilized.
Viscosity is the resistance of a fluid to shear. Capillary viscometers measure the time for a fixed volume of oil to flow by gravity through a calibrated tube at constant temperature, and kinematic viscosity equals flow time multiplied by the tube constant. Rotational instruments measure the torque needed to turn a spindle or cup in the sample at a known speed, giving dynamic viscosity. Stabinger-type cells combine a rotating tube with a floating inner rotor and a density measurement to give both dynamic and kinematic values. Cold cranking and mini-rotary tests apply defined shear or cooling profiles to simulate cold-start behavior, while tapered bearing simulators apply very high shear at 150 °C to measure HTHS viscosity. Precise temperature control is the dominant error source in every method.
A typical cold cranking simulator configuration offers temperature range of −35 to 0 °C, temperature stability of ± 0.05 °C, viscosity range of 500–10,000 mPa·s, shear rate of ≈ 10⁵–10⁶ s⁻¹, sample volume of ≈ 0.25–0.5 mL and rotor speed range of ≈ 50–300 rpm. Values are indicative of the product class; the exact figures for each configuration are confirmed on the quotation and the calibration documents supplied with the instrument.
The Cold Cranking Simulator is used in SAE J300 engine oil grading, additive and viscosity modifier development and passenger-car and diesel engine oil QC. Across the wider Viscosity Testing class, laboratories also deploy it for lubricant grading (ISO VG, SAE), engine oil qualification and base oil and additive blending.
Relevant standards and methods include ASTM D445, ISO 3104, ASTM D2270, ASTM D5293, ASTM D4684, ASTM D4683 and ASTM D7042. Certification candidates for this model are CE and ISO 9001 manufacturing; each is to be confirmed for the specific configuration before shipment, and certificates are supplied only where they have been issued.
Install viscosity equipment on a vibration-free bench away from drafts and direct sun. Calibrate viscometers with certified reference oils traceable to national standards, and verify thermometer or temperature probe accuracy at each test temperature. Glass capillaries must be cleaned thoroughly, since residue changes flow time, and inspected for chips. Replace circulating bath fluid on schedule and check stirring for uniformity. Record the capillary constants and requalify tubes periodically. Automatic units need periodic verification of sensors and pumps. Follow the standard's precision statement when deciding whether duplicate runs agree, and keep records of reference oil checks to support accreditation such as ISO/IEC 17025.
Choose a cold cranking simulator for engine oil blending or qualification against SAE J300 W grades. It is unnecessary for industrial oils tested only at 40 and 100 °C. Specify temperature range, coolant type, calibration oil set, software for grade evaluation, and sample-loading method.
Choose the variant from the viscosity parameter that the specification requires. Kinematic viscosity at 40 and 100 °C for industrial and engine oil grading points to a capillary bath, either manual for low volume or automatic for high throughput. Laboratories needing small samples, wide range and density in one run suit a Stabinger viscometer. Engine oil qualification to SAE J300 also needs cold cranking, mini-rotary and high-temperature high-shear instruments. Compare temperature stability, sample volume, number of tubes, cleaning approach and whether the instrument meets the method's dimensional and precision requirements. Consider operator time, since automatic units reduce cleaning and timing errors. For the RFQ, state the methods, temperature range, expected viscosity range, daily sample count and report format.
| Temperature range | −35 to 0 °C |
|---|---|
| Temperature stability | ± 0.05 °C |
| Viscosity range | 500–10,000 mPa·s |
| Shear rate | ≈ 10⁵–10⁶ s⁻¹ |
| Sample volume | ≈ 0.25–0.5 mL |
| Rotor speed range | ≈ 50–300 rpm |
| Measurement time | ≈ 2–5 min |
| Cooling system | Refrigerated circulator with coolant |
| Applicable method | ASTM D5293, ISO 15323 |
Indicative values for this product class. Final specifications are confirmed on the quotation and in the calibration documents supplied with the instrument.
It gives apparent viscosity at a low temperature and high shear, indicating whether an engine can crank fast enough to start in cold weather. It is one requirement of W-grade classification.
The measured rotor speed is converted to viscosity through a calibration curve from reference oils at each temperature. Without fresh, certified calibration oils, results can drift and fail the method's precision limits.
No. The two tests measure different properties: cold cranking gives apparent viscosity under high shear, while mini-rotary measures pumpability and yield stress at low shear. SAE J300 requires both for W grades.
Dynamic viscosity is the internal resistance to shear, measured in mPa·s. Kinematic viscosity is dynamic viscosity divided by density, measured in mm²/s. Capillary methods give kinematic values, while rotational methods give dynamic values.
Industrial and engine oils are commonly measured at 40 °C and 100 °C for grade assignment and viscosity index. Low-temperature methods run from about −40 to 0 °C, and HTHS viscosity is measured at 150 °C.
Viscosity index is calculated from kinematic viscosities at 40 °C and 100 °C using the procedure in ASTM D2270 or ISO 2909. A higher index means viscosity changes less with temperature.

手动运动黏度恒温浴及玻璃毛细管黏度计



