跳转到主要内容
Lab Manufacturers China

CO2 培养箱

CO2 Incubators

CON · 细胞培养设备 · 5 个型号

作者:Lab Manufacturers China 技术团队最近复核 适用标准:IEC 61010-1, IEC 61010-2-010, IEC 61326-1

参考标准

  • IEC 61010-1
  • IEC 61010-2-010
  • IEC 61326-1
  • EN 12469
  • ISO 14644-1
  • USP <1046>

什么是CO2 培养箱?

CO2 培养箱维持哺乳动物细胞培养所需的温度、湿度和二氧化碳浓度,使碳酸氢盐缓冲培养基保持约 pH 7.4,广泛用于细胞培养、干细胞、生物工艺和临床实验室的常规与受监管工作。

5 个型号 · CO2 培养箱

以下 5 个型号为当前配置,具体参数以报价单为准。

气套式 CO2 培养箱

Air-Jacketed CO2 Incubator

  • CON-0001
  • CO2 培养箱
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
腔体容积:
50–200 L (model dependent)
温度范围:
ambient +5 to 50 °C, set at 37 °C
温度均一性:
within ±0.3 °C at 37 °C
CO2 范围:
0–20%, controlled at 5%
CO2 准确度:
within ±0.1–0.3% (infrared sensor)
湿度:
above 90% relative humidity with water pan

气套式 CO2 培养箱以加热空气包围腔体,维持 37 °C 和 5% CO2,升温迅速且开门后恢复快,适合看重维护简便和轻便的常规细胞培养实验室。

索取报价
查看详情

水套式 CO2 培养箱

Water-Jacketed CO2 Incubator

  • CON-0002
  • CO2 培养箱
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
腔体容积:
150–300 L (model dependent)
温度范围:
ambient +5 to 50 °C, set at 37 °C
温度均一性:
within ±0.2 °C at 37 °C
水套容积:
approx. 20–40 L distilled water
CO2 范围:
0–20%, controlled at 5%
CO2 准确度:
within ±0.1–0.3% (infrared sensor)

水套式 CO2 培养箱以密封水套包围腔体,温度缓冲能力极强,可在断电和频繁开门时保持 37 °C,适合对培养稳定性要求苛刻的场景。

索取报价
查看详情

直热式 CO2 培养箱

Direct-Heat CO2 Incubator

  • CON-0003
  • CO2 培养箱
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
腔体容积:
50–180 L (model dependent)
温度范围:
ambient +5 to 50 °C, set at 37 °C
温度均一性:
within ±0.3–0.5 °C at 37 °C
CO2 范围:
0–20%, controlled at 5%
CO2 准确度:
within ±0.1–0.3%
湿度:
above 90% relative humidity with water pan

直热式 CO2 培养箱将加热元件置于腔壁上并依靠自然对流而非风扇,升温快、易清洁、培养空间内无运动部件,且采购成本较低。

索取报价
查看详情

三气低氧 CO2 培养箱

Triple-Gas Hypoxia CO2 Incubator

  • CON-0004
  • CO2 培养箱
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
氧浓度范围:
0.1–21%
氧控制准确度:
within ±0.2% at low set points
CO2 范围:
0–20%, controlled at 5%
温度范围:
ambient +5 to 50 °C, set at 37 °C
氧传感器:
Zirconia or electrochemical, calibration every 1–3 months
气体供应:
CO2 and nitrogen, or premixed triple gas

三气培养箱通过向腔内补入氮气同时控制氧气和二氧化碳,可将氧浓度维持在 0.1% 至 21%,用于对生理氧水平敏感的干细胞、胚胎和肿瘤培养。

索取报价
查看详情

振荡式 CO2 培养箱

CO2 Shaking Incubator

  • CON-0005
  • CO2 培养箱
  • CE (to be confirmed)
  • ISO 9001 manufacturing (to be confirmed)
  • IEC 61010-1 safety (to be confirmed)
振荡转速:
50–300 rpm, within ±1 rpm
振荡直径:
10–50 mm (model dependent)
温度范围:
ambient +5 to 60 °C, set at 37 °C
CO2 范围:
0–20%, controlled at 5%
容器容量:
up to 4 x 2 L flasks or 8 microplate positions
工作体积:
10 mL–2 L per vessel

振荡式 CO2 培养箱将轨道振荡与温度和二氧化碳控制结合,使悬浮细胞和微载体培养物在 37 °C、pH 7.4 条件下持续混匀,获得均匀的营养与氧接触。

索取报价
查看详情

对比CO2 培养箱型号

型号腔体容积温度范围温度均一性CO2 范围CO2 准确度
气套式 CO2 培养箱CON-000150–200 L (model dependent)ambient +5 to 50 °C, set at 37 °Cwithin ±0.3 °C at 37 °C0–20%, controlled at 5%within ±0.1–0.3% (infrared sensor)
水套式 CO2 培养箱CON-0002150–300 L (model dependent)ambient +5 to 50 °C, set at 37 °Cwithin ±0.2 °C at 37 °C0–20%, controlled at 5%within ±0.1–0.3% (infrared sensor)
直热式 CO2 培养箱CON-000350–180 L (model dependent)ambient +5 to 50 °C, set at 37 °Cwithin ±0.3–0.5 °C at 37 °C0–20%, controlled at 5%within ±0.1–0.3%
三气低氧 CO2 培养箱CON-0004100–200 Lambient +5 to 50 °C, set at 37 °C—0–20%, controlled at 5%—
振荡式 CO2 培养箱CON-0005—ambient +5 to 60 °C, set at 37 °C—0–20%, controlled at 5%—

以上数值代表该产品类别的常见水平。最终参数以报价单及随仪器提供的校准文件为准。

CO2 培养箱的工作原理是什么?培养中的细胞无法自行调节环境,因此培养箱提供培养基无法提供的条件。温度由腔体周围的水套、气套或直接加热元件配合 PID 控制和超温切断装置维持在 37 °C。二氧化碳由红外或热导传感器测量,并从气瓶注入以维持 5% CO2,使碳酸氢盐缓冲培养基处于生理 pH。水盘或加湿喷射使相对湿度保持在 90% 以上,防止蒸发导致盐和渗透物浓缩并改变 pH 设定点。铜质内壁或抗菌内胆板以及 HEPA 滤器可降低污染风险,内层玻璃门可在打开外门时减少气体交换。部分机型加入氮气以控制氧浓度,用于低氧培养。## CO2 培养箱有哪些型号?- 气套式 CO2 培养箱:气套式 CO2 培养箱以加热空气包围腔体,维持 37 °C 和 5% CO2,升温迅速且开门后恢复快,适合看重维护简便和轻便的常规细胞培养实验室。- 水套式 CO2 培养箱:水套式 CO2 培养箱以密封水套包围腔体,温度缓冲能力极强,可在断电和频繁开门时保持 37 °C,适合对培养稳定性要求苛刻的场景。- 直热式 CO2 培养箱:直热式 CO2 培养箱将加热元件置于腔壁上并依靠自然对流而非风扇,升温快、易清洁、培养空间内无运动部件,且采购成本较低。- 三气低氧 CO2 培养箱:三气培养箱通过向腔内补入氮气同时控制氧气和二氧化碳,可将氧浓度维持在 0.1% 至 21%,用于对生理氧水平敏感的干细胞、胚胎和肿瘤培养。- 振荡式 CO2 培养箱:振荡式 CO2 培养箱将轨道振荡与温度和二氧化碳控制结合,使悬浮细胞和微载体培养物在 37 °C、pH 7.4 条件下持续混匀,获得均匀的营养与氧接触。## 如何选择合适的型号?应首先选择腔体类型。水套式培养箱在断电期间温度最稳定,适合开门频繁的实验室;气套式和直热式升温更快、重量更轻且维护更少。其次是污染控制:铜质内壁、风路 HEPA 过滤以及高温或汽化过氧化氢消毒程序可减少繁忙实验室中的损失。还应检查 CO2 与温度准确度和均一性、开门后的恢复时间、湿度控制,以及干细胞或低氧实验是否需要氧浓度控制。若培养箱保存不可替代的培养物或属于受监管流程,应评估报警功能、数据记录、权限管理和远程监控。订购前应确认可叠放性、搁板可调性和可选腔体容积。## CO2 培养箱用于哪些领域?CO2 培养箱广泛用于常规哺乳动物细胞系培养、干细胞与原代细胞维持、疫苗与重组蛋白生产、临床细胞遗传学与病毒培养、低氧培养与类器官生长。## 适用哪些标准与认证?CO2 培养箱相关工作通常依据 IEC 61010-1、IEC 61010-2-010、IEC 61326-1、EN 12469、ISO 14644-1、USP <1046>。CE 等仪器认证需按具体配置逐项确认,不作默认承诺。## 采购前需要考虑哪些要点?培养箱应安装在水平台面,远离阳光直射、热源和气流,并为气路连接和通风预留空间。水盘应加无菌蒸馏水并每周更换,因为积水是常见的污染来源;部分机型采用加湿喷射,避免存水。安装时及至少每六个月一次,应使用经校准的外置仪器核查温度和 CO2 准确度,并记录开门后的恢复时间。搁板不应过载,因为空气循环是保持均一性的必要条件。溅洒应立即处理,只使用兼容的消毒剂,并在维修前按厂家程序进行消毒。应记录清洁、校准和消毒事件,以便追溯培养问题的原因。

关于CO2 培养箱的常见问题

水套式腔体在断电和频繁开门时温度更稳定,但重量更大且维护更多。气套式和直热式升温更快、恢复迅速且便于搬运,适合多数常规实验室。

需要协助选型CO2 培养箱?

索取报价